FDA Adverse Event Summary report: N

ENVISION G 2 SYSTEM/AP, RABBIT/MOUSE (PERMANENT RED)

MDR report key: 5544907 · Received April 4, 2016

Report

Report Number
9610099-2016-00002
Date Received
April 4, 2016
Date of Event
March 1, 2016
Report Date
April 3, 2017
Manufacturer
DAKO DENMARK A/S
Product Code
PPM
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

FINAL FOLLOW-UP INFORMATION: ROOT CAUSE THE CUSTOMER MISTAKENLY USED VISUALIZATION SYSTEM K5355, ENVISION G/2 SYSTEM/AP, RABBIT/MOUSE (PERMANENT RED), ON THE DAKO OMNIS INSTRUMENT; THE DETECTION SYSTEM IS ONLY VALIDATED FOR USE ON DAKO AUTOSTAINER INSTRUMENTS. THE CUSTOMER DETECTED THAT THE STAINING RESULT WAS NOT WITHIN SPECIFICATION AND FOUND OUT THAT THE WRONG INSTRUMENT WAS USED FOR STAINING WITH K5355. CORRECTIVE ACTIONS AS RESULT OF THE INVESTIGATION: NO CAPA HAS BEEN RAISED, SINCE THE CUSTOMER USED K5355 IN ERROR; THIS SYSTEM IS VALIDATED FOR USE ON THE DAKO AUTOSTAINER INSTRUMENTS AND IS NOT VALIDATED FOR USE ON THE DAKO OMNIS, WHICH LED TO THE WEAK STAINING.

Description of Event or Problem · 1

(B)(6) HOSPITAL COMPLAINED TO DAKO OF WEAK STAINING USING THE VISUALIZATION SYSTEM K5355, ENVISION G/2 SYSTEM/AP, RABBIT/MOUSE (PERMANENT RED), WHICH MAY HAVE LED TO FALSE NEGATIVE TEST RESULTS AND MAY HAVE CONTRIBUTED TO INCORRECT PATIENT DIAGNOSIS. THE HOSPITAL USED A DAKO ANTIBODY AGAINST MELAN A WITH THE K5355 PRODUCT (LOT 20015008), ON THE DAKO OMNIS PLATFORM. THE EXPIRATION DATE OF LOT 20015008 OF K5355 WAS 2015-10-31. THE ASSAY IN QUESTION WAS APPARENTLY RUN BY THE CUSTOMER IN 2016; IT IS POSSIBLE THAT EXPIRED PRODUCT WAS USED. DAKO IS IN THE PROCESS OF OBTAINING THE INSTRUMENT LOG FILES FROM THE CUSTOMER IN ORDER TO CONFIRM THIS POSSIBILITY. ACCORDING TO THE IFU FOR K5355, THE VISUALIZATION SYSTEM IS INTENDED FOR USE IN MANUAL STAINING PROCEDURES AND WITH THE DAKO AUTOSTAINER INSTRUMENTS. USE OF THE PRODUCT ON THE DAKO OMNIS HAS NOT BEEN VALIDATED. THE PRELIMINARY INFORMATION SUGGESTS THAT USE OF EXPIRED REAGENTS AND/OR NOT FOLLOWING THE IFU FOR THE PRODUCT WAS THE MOST LIKELY CAUSE OF THIS INCIDENT; HOWEVER ADDITIONAL INVESTIGATION CONTINUES TO BE CONDUCTED TO DETERMINE THE ROOT CAUSE. THE INVESTIGATION IS ONGOING TO DETERMINE THE ROOT CAUSE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
199971 ENVISION G 2 SYSTEM/AP, RABBIT/MOUSE (PERMANENT RED) ENVISION G/2 SYSTEM/AP, RABBIT/MOUSE (PERMANENT RED) PPM DAKO DENMARK A/S K5355 20015008

Patients

Seq Age Sex Outcome Treatment
1 Other