SILVERHAWK ATK
Report
- Report Number
- 2183870-2016-00294
- Event Type
- Malfunction
- Date Received
- April 4, 2016
- Date of Event
- March 1, 2016
- Report Date
- March 10, 2016
- Manufacturer
- COVIDIEN
- Product Code
- MCW
- PMA / PMN Number
- K061188
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THE PHYSICIAN INSERTED THE SILVERHAWK LX-M THROUGH THE 7F SHEATH WITH DIFFICULTY. FOUR PASSES WERE MADE IN THE SFA. THERE WAS DIFFICULTY REMOVING THE SILVERHAWK FROM THE SHEATH TO CLEAN. THE PHYSICIAN NOTICED A PIECE OF THE NOSE CONE PEELED AWAY NEAR THE CUTTER CHAMBER. AS A RESULT A NEW SILVERHAWK WAS OPENED TO COMPLETE THE PROCEDURE. NO INJURY REPORTED. EVALUATION OF THE RETURNED DEVICE WAS COMPLETED ON MARCH 28, 2016. A REVIEW OF THE MANUFACTURING RECORDS FOR THIS LOT DID NOT REVEAL ANY NON-CONFORMITY RELEVANT TO THIS REPORTED EVENT. THE REPORTED EVENT HAS BEEN CONFIRMED. THE SURFACE OF THE DISTAL ASSEMBLY BENEATH THE CUTTER WINDOW SHOWED THE MATERIAL WAS SLICED/SHAVED. THE STAINLESS STEEL COILS AT THE SEGMENT OF THE DISTAL ASSEMBLY WHERE THE DAMAGE WAS OBSERVED SHOWED SIGNS OF DEFORMATION. THE COILS WERE ELONGATED OR COMPRESSED AT AREAS AROUND/AT THE SHAVED SEGMENT. THE ROOT CAUSE OF THE DAMAGE ENCOUNTERED COULD NOT BE DETERMINED. POSSIBLE CONTRIBUTING FACTORS SUCH AS; LESION MORPHOLOGY, VESSEL ANATOMY, SHEATH INTERACTION, AND PROCEDURAL TECHNIQUE COULD NOT BE EVALUATED IN THIS INVESTIGATION, HOWEVER, ACCORDING TO THE PRODUCT LABELING FOR THE SILVERHAWK LX-M THE RECOMMENDED SHEATH SIZE IS 8F. THE INSTRUCTIONS FOR USES INCLUDES THE FOLLOWING: USE OF SHEATHS SMALLER THAN THOSE RECOMMENDED MAY COMPROMISE DEVICE PERFORMANCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 200916 | SILVERHAWK ATK | CATHETER, PERIPHERAL, ATHERECTOMY | MCW | COVIDIEN | P4055 | A111789 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00055 YR |