FDA Adverse Event Injury Summary report: N

NIPRO ELISIO-H HEMODIALYZER

MDR report key: 5543839 · Received April 4, 2016

Report

Report Number
9610987-2016-00001
Event Type
Injury
Date Received
April 4, 2016
Date of Event
January 4, 2016
Report Date
April 1, 2016
Manufacturer
NIPRO CORPORATION (ODATE)
Product Code
KDI
PMA / PMN Number
K131935
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

.

Description of Event or Problem · 1

AFTER 2 HOURS AND 30 MINUTES OF INITIATION OF DIALYSIS TREATMENT, STAFF NOTICED BLOOD WAS LEAKING BETWEEN THE ARTERIAL HEADER AND CASE OF THE DIALYZER. THERE WAS NO LEAKAGE DURING PRIMING. DIALYSIS SESSION WAS STOPPED RIGHT AWAY, AND BLOOD WAS NOT RETURNED. ACCORDING TO THE PHYSICIAN, THE PATIENT LOST ABOUT 400ML OF BLOOD. PATIENT REQUIRED 500ML OF PLASMA RIGHT AWAY. PATIENT'S DIALYSIS ORDER: ON-LIN POST-DILUTION HDF EXPECTED TIME: FOUR (4) HOURS. PARAMETERS 20 MINUTES BEFORE DISCONNECTING PATIENT: BLOOD FLOW RATE: THREE-HUNDRED FIFTY (350) ML/MIN, DIALYSATE FLOWRATE: 600ML/MIN, ARTERIAL PRESSURE: ONE-HUNDRED SEVENTY-ONE (171) MMHG, VENOUS PRESSURE: ONE-HUNDRED THIRTY-NINE (139) MMHG, QFI/QB = 31%(RATIO OF REINFUSION), TOTAL SUBSTITUTION VOLUME: THIRTEEN (13) LITERS, UF/HOURS: 950ML/HOUR, TOTAL UF LOST: TWO-THOUSAND ONE-HUNDRED SEVENTY (2170) ML (PROGRAMMING MADE ON 4 HOURS AND 3.8 LITERS OF WEIGHT LOSS). OTHER MEDICAL EQUIPMENT USED: GAMBRO ARTHIS PHYSIO BLOODLINES, NIPRO AV1525 HTC 15R/AV 1525 TC 15B BLOOD-NEEDLES, ARTHIS PHYSIO DIALYSIS MACHINE PRIMING CONDITIONS FOR DIALYSIS MACHINE: FLOWRATE BETWEEN 100 AND 400 ML/MIN, WITH AN AVERAGE FLOWRATE OF 150ML/MIN (TIME OF PRIMING = 9.3 MINUTES) TOTAL VOLUME: 1400ML AND ANOTHER 300ML BEFORE CONNECTING THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
201329 NIPRO ELISIO-H HEMODIALYZER DIALYZER KDI NIPRO CORPORATION (ODATE) DD+ELISIO-21H 15J01D

Patients

Seq Age Sex Outcome Treatment
1 78 YR Required Intervention