FDA Adverse Event Malfunction Summary report: N

CNS-6201A

MDR report key: 5543024 · Received April 1, 2016

Report

Report Number
8030229-2016-00138
Event Type
Malfunction
Date Received
April 1, 2016
Date of Event
March 2, 2016
Report Date
March 2, 2016
Manufacturer
NIHON KOHDEN TOMIOKA CORPORATION
Product Code
MHX
PMA / PMN Number
K102376
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER REPORTED THAT THE CNS (CENTRAL MONITORING SYSTEM) IS GETTING A SSD ACCESS ERROR. THE CUSTOMER SENT THE LOG FILES TO NIHON KOHDEN TECHNICAL SUPPORT FOR ANALYSIS. A LOANER UNIT WAS SENT OUT TO THE CUSTOMER WHILE THIS REPORT IS UNDER INVESTIGATION. NIHON KOHDEN WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

THE CUSTOMER REPORTED THAT THE CENTRAL NURSE'S STATION (CNS) DISPLAYED A SSD ACCESS ERROR. THE CUSTOMER SENT THE LOG FILES AND THE DEVICE TO NIHON KOHDEN FOR ANALYSIS, EVALUATION, AND REPAIR. THE SSD ACCESS ERROR COULD NOT BE DUPLICATED. WHEN BOOTING UP, THE DEVICE DISPLAYED A "CNS6201.EXE- CORRUPT FILE" ERROR MESSAGE, INDICATING THAT THE FILE OR DIRECTORY C:\$MFT WAS CORRUPT AND UNREADABLE. RUNNING A SCAN DISK RESOLVED THE ISSUE AND THE CNS STARTED TO FUNCTION NORMALLY. NIHON KOHDEN'S REPAIR CENTER COMPLETED ALL OF THE STEPS IN THE MAINTENANCE CHECK SHEET OF THE SERVICE MANUAL AND PERFORMED AN EXTENDED TEST. THE REPAIRED DEVICE WAS RETURNED TO THE CUSTOMER. NIHON KOHDEN WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE CNS (CENTRAL MONITORING SYSTEM) IS GETTING A SSD ACCESS ERROR. THE CUSTOMER SENT THE LOG FILES TO NIHON KOHDEN TECHNICAL SUPPORT FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
197190 CNS-6201A CENTRAL MONITOR SYSTEM MHX NIHON KOHDEN TOMIOKA CORPORATION CNS-6201A

Patients

Seq Age Sex Outcome Treatment
1