FDA Adverse Event
Other
Summary report: N
MEDISORB MULTI ABSORBER
MDR report key: 554177
·
Received April 19, 2004
Report
- Report Number
- 9610105-2004-00007
- Event Type
- Other
- Date Received
- April 19, 2004
- Date of Event
- March 31, 2004
- Report Date
- April 1, 2004
- Manufacturer
- INSTRUMENTARIUM CORP. DATEX-OHMEDA DIV
- Product Code
- BSF
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
CUSTOMER COMPLAINT: CANNOT VENTILATE PT, NOR IN MANUAL NOR IN VENT. MODE. THE FIELD ENGINEER DISCOVERED THAT A NURSE HAD MOUNTED A NEW SODALIME CANISTER JUST BEFORE THE START OF A NEW PT AND THAT THE CANISTER STILL HAD THE GREEN COVER-LIP ON. CO TESTED SEVERAL CANISTERS AND IT IS INDEED EASY TO INSTALL A CANISTER WITH THE COVER LIP STILL ON. WHEN YOU DO SO AND CARRY OUT A NORMAL FUNCTIONAL TEST THE AVANCE WILL WARN YOU, BUT SOMEONE DURING AN OPERATION HAS TO REPLACE A CANISTER IT CAN EASILY HAPPEN AGAIN. EXTRA: THREEE WEEKS AGO THE SAME HOSP MANAGED TO PUT THAT CANISTER THE OTHER WAY AROUND (DRAIN RESERVOIR TO THE FRONT).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MEDISORB MULTI ABSORBER | DATEX-OHMEDA MULTI ABSORBER, DISPOSABLE | BSF | INSTRUMENTARIUM CORP. DATEX-OHMEDA DIV | 8003138 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR | Other |