FDA Adverse Event Other Summary report: N

MEDISORB MULTI ABSORBER

MDR report key: 554177 · Received April 19, 2004

Report

Report Number
9610105-2004-00007
Event Type
Other
Date Received
April 19, 2004
Date of Event
March 31, 2004
Report Date
April 1, 2004
Manufacturer
INSTRUMENTARIUM CORP. DATEX-OHMEDA DIV
Product Code
BSF
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

CUSTOMER COMPLAINT: CANNOT VENTILATE PT, NOR IN MANUAL NOR IN VENT. MODE. THE FIELD ENGINEER DISCOVERED THAT A NURSE HAD MOUNTED A NEW SODALIME CANISTER JUST BEFORE THE START OF A NEW PT AND THAT THE CANISTER STILL HAD THE GREEN COVER-LIP ON. CO TESTED SEVERAL CANISTERS AND IT IS INDEED EASY TO INSTALL A CANISTER WITH THE COVER LIP STILL ON. WHEN YOU DO SO AND CARRY OUT A NORMAL FUNCTIONAL TEST THE AVANCE WILL WARN YOU, BUT SOMEONE DURING AN OPERATION HAS TO REPLACE A CANISTER IT CAN EASILY HAPPEN AGAIN. EXTRA: THREEE WEEKS AGO THE SAME HOSP MANAGED TO PUT THAT CANISTER THE OTHER WAY AROUND (DRAIN RESERVOIR TO THE FRONT).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MEDISORB MULTI ABSORBER DATEX-OHMEDA MULTI ABSORBER, DISPOSABLE BSF INSTRUMENTARIUM CORP. DATEX-OHMEDA DIV 8003138 *

Patients

Seq Age Sex Outcome Treatment
1 43 YR Other