COULTER LH 750 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2016-00314
- Event Type
- Malfunction
- Date Received
- March 31, 2016
- Date of Event
- March 10, 2016
- Report Date
- March 10, 2016
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K061574
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FIELD SERVICE ENGINEER (FSE) EVALUATED THE INSTRUMENT ON (B)(4) 2016. THE FSE FOUND THAT THE TUBING AT PINCH VALVE VL53B WAS DEFECTIVE. THE FSE REPLACED THE TUBING, WHICH REPAIRED THE LEAK. THE REPAIRS WERE VERIFIED BY THE FSE. (B)(4).
THE INSTRUMENT MODEL WAS CHANGED FROM LH500 TO LH750. THE 510(K) NUMBER WAS CHANGED FROM K042724 TO K061574.
THE CUSTOMER REPORTED A BLUE LEAK FROM THE COULTER LH 500 HEMATOLOGY ANALYZER. THE VOLUME OF THE LEAK WAS A FEW DROPS AND WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF LABORATORY COAT AND GLOVES AT THE TIME OF THE EVENT. THERE WAS NO REPORT OF INJURY OR BIOHAZARD EXPOSURE TO OPEN WOUNDS OR MUCOUS MEMBRANES. ERRONEOUS PATIENT RESULTS WERE NOT GENERATED AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT.
THE CUSTOMER REPORTED A BLUE LEAK FROM THE COULTER LH 750 HEMATOLOGY ANALYZER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 194374 | COULTER LH 750 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |