FDA Adverse Event Malfunction Summary report: N

HAKIM INLINE PROGRAMMABLE VALVE SG

MDR report key: 5538688 · Received March 31, 2016

Report

Report Number
1226348-2016-10249
Event Type
Malfunction
Date Received
March 31, 2016
Manufacturer
CODMAN & SHURTLEFF, INC. / MEDOS S.A.
Product Code
JXG
PMA / PMN Number
PK992173
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). UPON COMPLETION OF THE INVESTIGATION IT WAS NOTED THAT THE IMAGES WERE TAKEN OF THE ¿AS RECEIVED¿ VALVE. THE POSITION OF THE CAM WHEN VALVE WAS RECEIVED WAS 30MMH2O. THE VALVE WAS HYDRATED FOR ABOUT 26 HOURS. THE VALVE WAS VISUALLY INSPECTED: NO DEFECTS WERE NOTED. THE VALVE WAS TESTED FOR PROGRAMMING WITH PROGRAMMER 82-3126 WITH SERIAL NUMBER (B)(4), THE VALVE PASSED THE TEST. THE VALVE WAS FLUSHED AND THE VALVE PASSED THE TEST NO OCCLUSION WAS NOTED. THE SIPHON GUARD WAS TESTED. THE VALVE PASSED THE TEST. THE VALVE WAS REFLUX TESTED. THE VALVE PASSED THE TEST. THE VALVE WAS LEAK TESTED, NO LEAKS WERE FOUND. THE SIPHON GUARD WAS REMOVED. THE VALVE WAS DRIED. THE VALVE WAS THEN PRESSURE TESTED, THE VALVE PASSED THE TEST. REVIEW OF THE HISTORY DEVICE RECORDS CONFIRMED THE VALVE PRODUCT CODE 82-3162, WITH LOT CTBCVB, CONFORMED TO THE SPECIFICATIONS WHEN RELEASED TO STOCK IN 19TH FEBRUARY 2015. NO ROOT CAUSE COULD BE DETERMINED, AS THE PROBLEM REPORTED BY THE CUSTOMER COULD NOT BE DUPLICATED. BASED ON THE RESULTS OF THIS INVESTIGATION NO FURTHER ACTION IS REQUIRED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR COMPLAINTS. AT THE PRESENT TIME THIS COMPLAINT IS CLOSED.

Additional Manufacturer Narrative · 1

(B)(4). UPON COMPLETION OF THE INVESTIGATION A FOLLOW UP REPORT WILL BE FILED.

Description of Event or Problem · 1

REP PRESENTED VERBALLY: BEFORE IMPLANTATION, SURGEON WAS UNABLE TO PRIME 823162. THEY USED A REPLACEMENT WITH NO DIFFICULTIES, DELAYING SURGERY BY SEVERAL MINUTES. SENDING VALVE IN FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
192740 HAKIM INLINE PROGRAMMABLE VALVE SG SHUNT, CENTRAL NERVOUS SYSTEM & COMPS JXG CODMAN & SHURTLEFF, INC. / MEDOS S.A. CTBCVB

Patients

Seq Age Sex Outcome Treatment
1