HEARTSINE SAMARITAN 300P AND PAD PAK
Report
- Report Number
- 3004123209-2016-00350
- Event Type
- Malfunction
- Date Received
- March 31, 2016
- Date of Event
- March 14, 2016
- Report Date
- May 12, 2016
- Manufacturer
- HEARTSINE TECHNOLOGIES LTD
- Product Code
- MKJ
- PMA / PMN Number
- K041067
- Removal / Correction Number
- Z-0124-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- PATIENT
Narratives
EXEMPTION NUMBER E2015058 HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER). H3 OTHER TEXT : NOT YET RETURNED TO MANUFACTURER
EXEMPTION NUMBER E2015058. HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER). THE HISTORY LOG FOR THIS DEVICE SHOWED THAT THE PAD-PAK WAS FIRST INSTALLED ON THE (B)(6) 2009 AND THAT IT HAD PERFORMED TO SPECIFICATION UP TO THE (B)(6) 2012. DURING THIS TIME THE DEVICE RECORDS MULTIPLE OCCASIONS IN WHICH THE TEMPERATURE WAS BELOW THE MINIMUM RECOMMENDED STAND-BY TEMPERATURE OF 0 DEGREES CELSIUS WITH A LOW OF -7.6 DEGREES CELSIUS RECORDED. ON THE (B)(6) 2012 THE DEVICE FAILED THE WEEKLY AUTO SELF-TEST DUE TO A LOW BATTERY. MULTIPLE MANUAL POWER UPS, MAINLY OF TEN MINUTES DURATION WERE RECORDED BETWEEN THE (B)(6) 2016 AND THE LAST LOG ENTRY ON THE (B)(6) 2016. INVESTIGATION REPORTED THE FAULT TO BE CONTRIBUTED TO MEMBRANE FAILURE. THE TEN MINUTE TIME OUTS WOULD SUGGEST A FAILING MEMBRANE. THE PAD-PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE, WHICH IS WHY THE "UNKNOWN" BOX HAS BEEN CHECKED IN THIS REPORT.
THERE WAS NO PATIENT INVOLVED IN THIS EVENT. UNIT POWERS ON BY ITSELF WITHOUT MANUAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 193606 | HEARTSINE SAMARITAN 300P AND PAD PAK | AUTOMATED EXTERNAL DEFIBRILLATOR | MKJ | HEARTSINE TECHNOLOGIES LTD | SAM | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |