FDA Adverse Event Other Summary report: N

GIRAFFE INCUBATOR

MDR report key: 553587 · Received November 4, 2004

Report

Report Number
1121732-2004-00017
Event Type
Other
Date Received
November 4, 2004
Date of Event
October 1, 2004
Report Date
November 4, 2004
Manufacturer
OHMEDA MEDICAL
Product Code
FMT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

IT WAS REPORTED THAT SMOKE WAS OBSERVED FROM THE NORTH WALL INTERFACE OF A GIRAFFE INCUBATOR AFTER THE HUMIDITY SETTINGS HAD BEEN CHANGED. AN INFANT WAS BEING TREATED IN THE UNIT WHEN THE EVENT OCCURRED. THE INFANT WAS REMOVED FROM THE UNIT AND TRANSFERRED TO ANOTHER INCUBATOR. THE UNIT WAS TAKEN OUT OF SERVICE AND QUARANTINED AT THE HOSPITAL. THE FIELD SERVICE ENGINEER REPLACED THE RELAY BOARD AND THE UNIT WAS PUT BACK INTO SERVICE THE SAME DAY. NO PT INJURY WAS REPORTED. PRIOR TESTING BY OHMEDA MEDICAL HAS DEMONSTRATED THAT SMOKE CANNOT ENTER THE INFANT COMPARTMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GIRAFFE INCUBATOR INFANT THERMOREGULATION FMT OHMEDA MEDICAL * *

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other