FDA Adverse Event Malfunction Summary report: N

CORTISOL

MDR report key: 5535832 · Received March 30, 2016

Report

Report Number
1823260-2016-00380
Event Type
Malfunction
Date Received
March 30, 2016
Date of Event
March 12, 2016
Report Date
April 26, 2016
Manufacturer
ROCHE DIAGNOSTICS
Product Code
NHG
PMA / PMN Number
K070788
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED IN (B)(6).

Additional Manufacturer Narrative · 1

A SPECIFIC ROOT CAUSE COULD NOT BE IDENTIFIED. ADDITIONAL INFORMATION FOR FURTHER INVESTIGATION WAS REQUESTED BUT WAS NOT PROVIDED.BASED ON A REVIEW OF QUALITY CONTROL DATA, A GENERAL REAGENT ISSUE CAN BE EXCLUDED.

Description of Event or Problem · 1

THE CUSTOMER COMPLAINED OF ERRONEOUS RESULTS FOR 1 PATIENT TESTED WITH BOTH CORTISOL AND CORTISOL II. THE CUSTOMER IS NOTICING A DIFFERENCE IN THE RESULTS BETWEEN THE 2 ASSAYS AND DOES NOT KNOW WHICH RESULT IS CORRECT. THE RESULTS FOR BOTH ASSAYS WERE REPORTED OUTSIDE OF THE LABORATORY. THE INITIAL CORTISOL RESULT WAS 5.87 NMOL/L. THE INITIAL CORTISOL II RESULT WAS 115.3 NMOL/L. THESE RESULTS WERE RUN ON E602 ANALYZER WITH SERIAL NUMBER (B)(4). THE SAMPLE WAS REPEATED AND THE CORTISOL RESULT WAS 4.80 NMOL/L. THE CORTISOL II RESULT WAS 109.7. THE CORTISOL RESULT WAS REPORTED OUTSIDE OF THE LABORATORY AS 5 NMOL/L AND THE CORTISOL II RESULT WAS REPORTED OUTSIDE OF THE LABORATORY AS 110 NMOL/L. THESE RESULTS WERE RUN ON E602 ANALYZER WITH SERIAL NUMBER (B)(4). THE CUSTOMER ALSO PROVIDED RESULTS FROM A PREVIOUS SAMPLE FROM THIS PATIENT OBTAINED ON AN UNKNOWN DATE AND THE CORTISOL RESULT WAS 9.0 NMOL/L. THE CORTISOL II RESULT WAS 114 NMOL/L. THE CUSTOMER BELIEVES THE ERRONEOUS RESULTS ARE DUE TO A PATIENT ANOMALY. NO ADVERSE EVENT OCCURRED. THE CORTISOL II ASSAY IS NOT SOLD IN THE UNITED STATES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
192192 CORTISOL ENZYME IMMUNOASSAY, CORTISOL, SALIVARY NHG ROCHE DIAGNOSTICS NA 18538001

Patients

Seq Age Sex Outcome Treatment
1