FDA Adverse Event Injury Summary report: N

FMP HIP

MDR report key: 5531713 · Received March 29, 2016

Report

Report Number
1644408-2016-00219
Event Type
Injury
Date Received
March 29, 2016
Date of Event
March 4, 2016
Report Date
March 4, 2016
Manufacturer
ENCORE MEDICAL, L.P.
Product Code
OQH
UDI-DI
00888912099936
PMA / PMN Number
K130365
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REASON FOR THIS REVISION SURGERY WAS HIP DISLOCATION. THE LENGTH OF THE IN-VIVO SERVICE WAS 2.4 MONTHS. THERE IS NO INFORMATION IN THIS COMPLAINT ABOUT ANY PATIENT INJURIES, ACTIVITIES, OR ACCIDENTS THAT MAY HAVE CONTRIBUTED TO THE NEED FOR THIS REVISION SURGERY. THE HEALTHCARE PROFESSIONAL INDICATED THERE WAS A SERIOUS RISK TO THE PATIENT. THERE WAS NO DELAY IN SURGERY AND ANOTHER SUITABLE DEVICE WAS AVAILABLE FOR USE. THE REVISION SURGERY WAS COMPLETED AS INTENDED. THE DEVICE WAS DISPOSED OF AT THE HOSPITAL AND NOT MADE AVAILABLE TO (B)(4) FOR EXAMINATION. A REVIEW OF THE DEVICE HISTORY RECORDS (DHRS) REVEALED NO DISCREPANCIES OR ISSUES WITH THE MANUFACTURING HISTORY OF THIS PART. ALL PARTS WERE FOUND TO MEET DESIGN AND MANUFACTURING SPECIFICATIONS. NO NON-CONFORMING MATERIAL REPORTS (NCMR'S) WERE ASSOCIATED WITH THIS PART. THE PRODUCT COMPLAINT REPORT HISTORY WAS REVIEWED AND NO TRENDS OR ON-GOING ISSUES WERE DEEMED AS PRESENT OR IN NEED OF REVIEW. THIS EVENT IS DEEMED TO BE NON-PRODUCT RELATED. THE SURGEON REPORTED NO ISSUES ASSOCIATED WITH THE EXPLANTED PRODUCT AND PROVIDED NO INFORMATION THAT DEFINITIVELY DESCRIBED THE ROOT CAUSE OR REASON OF THE DISLOCATION. THE SCOPE OF THIS INVESTIGATION IS LIMITED WITHOUT HAVING THE PARTS AVAILABLE TO (B)(4)FOR EVALUATION. OTHER CONDITIONS RELATING TO THIS EVENT COULD NOT BE DETERMINED WITH CONFIDENCE. FACTORS THAT MAY CONTRIBUTE TO A DISLOCATION NOT ASSOCIATED WITH THE IMPLANT ARE: IMPROPER IMPLANT SELECTION, DEGENERATIVE BONE DISEASE, AND IMPROPER SURGICAL TECHNIQUE. INVENTORY CONTAINMENT IS NOT REQUIRED AS THERE ARE NO INDICATIONS OF A PRODUCT OR PROCESS ISSUE AFFECTING IMPLANT SAFETY OR EFFECTIVENESS.

Description of Event or Problem · 1

REVISION SURGERY - DUE TO THE PATIENT DISLOCATING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
189081 FMP HIP LINER/NON-HOODED-NEU, MP10, HXE-PLUS, 44MM OQH ENCORE MEDICAL, L.P. 764N1002 00888912099936

Patients

Seq Age Sex Outcome Treatment
1 74 YR Other| R