MDR report key: 5528664 · Received March 28, 2016

Report

Report Number
9611253-2016-00012
Event Type
Malfunction
Date Received
March 28, 2016
Date of Event
January 21, 2016
Report Date
October 7, 2016
Manufacturer
NAKANISHI INC.
Product Code
EGS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURED IN (B)(4), BUT THE SIMILAR PRODUCTS ARE MARKETED IN THE U.S. UNDER K962540.

Additional Manufacturer Narrative · 1

NAKANISHI SENT TWO EMAIL MESSAGES TO (B)(6) ON APRIL 1, 2016 AND APRIL 29, 2016 TO REQUEST PATIENT'S ID AND WEIGHT THAT WERE MISSING IN THE INITIAL REPORT.NAKANISHI RECEIVED AN EMAIL MESSAGE FROM (B)(6) ON MAY 3, 2016 THAT STATES (B)(6) HAD NO ADDITIONAL PATIENT INFORMATION.

Additional Manufacturer Narrative · 1

UPON RECEIVING THE DEVICE INVOLVED IN THE MDR EVENT, NAKANISHI CONDUCTED A FAILURE ANALYSIS OF THE RETURNED DEVICE: METHODOLOGY USED: A) NAKANISHI EXAMINED THE DEVICE HISTORY RECORD FOR THE SUBJECT BA105 DEVICE (SERIAL NUMBER (B)(4)). THERE WERE NO PROBLEMS OBSERVED DURING THE MANUFACTURING OR TESTING NOTED IN THE DHR. B) NAKANISHI CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE AND PERFORMED A SIMPLE MOVEMENT TEST. C) ROTATION CHECK - NAKANISHI OPERATED THE MOTOR 40,000RPM AND CHECKED THE ROTATION. NAKANISHI COULD NOT FIND ANY ABNORMAL SOUND OR ABNORMAL VIBRATION. D) CHUCK PART CHECK - NAKANISHI REMOVED THE HEAD CAP AND OBSERVED THE CHUCK PART. NAKANISHI COULD NOT DISCERN ANY DAMAGE THAT COULD HAVE LEAD TO CHUCK PART FAILURE. - WHEN NAKANISHI MOUNTED A BUR, NAKANISHI CONFIRMED THAT THE BUR WAS CHUCKED NORMALLY. (NAKANISHI TRIED TO PULL OUT THE BUR WITH FINGERS, BUT WAS UNABLE TO DO SO.) - COMPARING THE BUR INSERTION IN THE RETURNED HANDPIECE TO A NEW HANDPIECE, NAKANISHI COULD NOT FIND ANY DIFFERENCE IN THE CHUCKING STATE. E) BUR RETENTION FORCE - NAKANISHI MOUNTED THE BUR INTO THE CHUCK FOLLOWING THE OPERATION MANUAL AND CHECKED THE RETENTION FORCE. - THE CHUCKING FORCE SPECIFICATION IS 45 N, BUT NAKANISHI MEASURED UP TO 95 N OF FORCE; THE BUR STILL COULD NOT BE PULLED OUT. F) NAKANISHI COULD NOT INVESTIGATE THE FALLEN BUR BECAUSE IT WAS NOT RETURNED TO NAKANISHI. ACCORDING TO THE DISTRIBUTOR, THE BUR THAT WAS USED MAY HAVE BEEN A HENRY SCHEIN BUR. THEREFORE, NAKANISHI IS NOT SURE IF THE DIMENSION OF THE BUR SHANK IS COMPLIANT WITH THE ISO STANDARD. G) VISUAL INSPECTION OF OTHER PARTS - NAKANISHI DISASSEMBLED THE HANDPIECE AND PERFORMED A VISUAL INSPECTION OF THE OTHER PARTS. - NAKANISHI COULD NOT FIND ANY DEFECTIVE PORTION SUCH AS DAMAGE OR ABNORMAL WEAR. H) LABELING REVIEW - NAKANISHI CONFIRMED THAT THE OPERATION MANUAL CONTAINS THE FOLLOWING VERBIAGE : "CAUTION: BEFORE USE, MAKE SURE THAT A BUR IS SECURELY LOCKED, PULLING IT GENTLY WITH FINGERS. MAKE A TRIAL RUN" I) CONCLUSION REACHED BASED ON THE INVESTIGATION AND ANALYSIS RESULT - NAKANISHI COULD NOT FIND ANY ABNORMAL PARTS DURING THE ABOVE INVESTIGATION. NAKANISHI DETERMINED THAT THE RETURNED PRODUCT WAS NORMAL. - NAKANISHI CONFIRMED THROUGH THE INVESTIGATION THAT THE SUFFICIENT RETENTION FORCE CAN BE OBTAINED IF A BUR THAT IS COMPLIANT WITH THE ISO STANDARD IS CHUCKED NORMALLY. NAKANISHI DETERMINED THAT A BUR CAN NOT BE PULLED OUT DUE TO THE USAGE LOAD. - NAKANISHI CONCLUDED THAT THE BUR HAD BEEN PULLED OUT DUE TO DRILLING LOAD WHEN THE BUR WAS NOT CHUCKED CORRECTLY.

Description of Event or Problem · 1

NAKANISHI RECEIVED AN EMAIL FROM (B)(6) ON MARCH 7, 2016 THAT STATES THE DENTIST WAS PERFORMING A TOOTH RESTORATION ON A PATIENT WHEN THE BUR FELL OUT OF THE HANDPIECE MID-TREATMENT. THE BUR WAS RETRIEVED FROM THE PATIENT'S MOUTH BY THE DENTIST AND THE PATIENT WAS NOT HURT AT ALL. ACCORDING TO (B)(6), THE BUR BEING USED WAS POSSIBLY A HENRY SCHEIN BUR ROUND STEEL (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
186577 BA105 HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL EGS NAKANISHI INC. BA105

Patients

Seq Age Sex Outcome Treatment
1 40 YR Other