FDA Adverse Event
Other
Summary report: N
REVEAL XL
MDR report key: 552414
·
Received April 19, 2004
Report
- Report Number
- 1034973-2004-00001
- Event Type
- Other
- Date Received
- April 19, 2004
- Date of Event
- February 11, 2004
- Report Date
- March 17, 2004
- Manufacturer
- CTI PET SYSTEMS, INC.
- Product Code
- KPS
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE END USER PERFORMED A 3D NORMALIZATION ON THE PET SCANNER. WHEN PROMPTED FOR THE CE-68 PHANTOM, THE END USER SELECTED 2D INSTEAD OF 3D. THIS AFFECTED THE STANDARD UPTAKE VALUES (SUV) OF 39 PTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | REVEAL XL | POSITRON EMISSION TOMOGRAPHY (PET) SCANNER | KPS | CTI PET SYSTEMS, INC. | 3600211-00 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN |