FDA Adverse Event Other Summary report: N

BD PARADIGM LINK BLOOD GLUCOSE MONITOR

MDR report key: 552061 · Received April 22, 2004

Report

Report Number
2243072-2004-00205
Event Type
Other
Date Received
April 22, 2004
Date of Event
March 23, 2004
Report Date
April 21, 2004
Manufacturer
BECTON DICKINSON
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

CONSUMER CALLED TO REPORT THEY HAVE BEEN GETTING VERY ERRATIC READINGS WITH THEIR PLINK METER. ANALYSIS RUN ON CLARK ERROR GRID USING MEAN AVG READING (60) AS REF POINT VS BD READING (74, 46) YIELDED DATA POINTS IN THE DE RANGE OF THE GRID, OUTSIDE ACCEPTABLE LIMITS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BD PARADIGM LINK BLOOD GLUCOSE MONITOR BD TEST STRIPS NBW BECTON DICKINSON NA 3064300

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other