FDA Adverse Event Injury Summary report: N

ZENITH COOK, INC.

MDR report key: 551869 · Received October 22, 2004

Report

Report Number
MW1033661
Event Type
Injury
Date Received
October 22, 2004
Manufacturer
COOK, INC.
Product Code
MIH
Adverse Event
Yes
Report Source
Voluntary report
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

ZENITH ENDOGRAFT PLACED IN 2003 WITHOUT INCIDENT. AT ONE YEAR FOLLOW UP, CONTRALATERAL LIMB HAD MIGRATED CAUDALLY ONE CM AT OVERLAP WITH MAIN DEVICE. THIS WAS NOT EVIDENT ON SIX MONTH F/U FILMS. PT WAS TREATED WITH ESLE 12-55 LIMB EXTENSION TO INCREASE THE DEVICE OVERLAP, IN 2004. PT HAD NO COMPLICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ZENITH COOK, INC. AAA STENT GRAFT MIH COOK, INC. * 1260957
2 ZENITH COOK INC. AAA STENT GRAFT MIH COOK INC * 1240587
3 ZENITH COOK INC AAA STENT GRAFT MIH COOK INC * 1233606

Patients

Seq Age Sex Outcome Treatment
1 71 YR Required Intervention