FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE SYSTEM

MDR report key: 5514648 · Received March 21, 2016

Report

Report Number
3004753838-2016-20937
Event Type
Malfunction
Date Received
March 21, 2016
Date of Event
February 5, 2016
Report Date
February 25, 2016
Manufacturer
DEXCOM, INC.
Product Code
OYC
UDI-DI
00386270000019
PMA / PMN Number
P130007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

PATIENT CONTACTED DEXCOM ON (B)(6) 2016 TO REPORT CONTINUOUS GLUCOSE MONITORING (CGM) INACCURACIES COMPARED TO BLOOD GLUCOSE (BG) METER THAT OCCURRED ON (B)(6) 2016. SENSOR WAS INSERTED ON (B)(6) 2016. AT THE TIME OF CONTACT, NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Description of Event or Problem · 1

SUBSEQUENT TO THE INITIAL MDR, THE COMPLAINT SENSOR WAS NOT RETURNED TO DEXCOM. THE TRANSMITTER (PART NUMBER STT-GL-003/SERIAL NUMBER (B)(4)/LOT NUMBER 5203772) WAS RETURNED ON 04/27/2016. THE RETURNED TRANSMITTER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THERE WAS NO FAILURE DETECTED RELATED TO THE COMPLAINT. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION. THE REPORTED EVENT OF INACCURACIES WAS NOT CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
170729 ANIMAS VIBE SYSTEM OYC OYC DEXCOM, INC. 9500-27 5205374 00386270000019

Patients

Seq Age Sex Outcome Treatment
1 44 YR