FDA Adverse Event Injury Summary report: N

SYSMEX CA-1500 CP

MDR report key: 5513971 · Received March 21, 2016

Report

Report Number
1000515253-2016-00007
Event Type
Injury
Date Received
March 21, 2016
Date of Event
February 24, 2016
Report Date
March 11, 2016
Manufacturer
SYSMEX CORPORATION
Product Code
GKP
PMA / PMN Number
K993299
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

AN OPERATOR IN (B)(6) WAS INJURED WHILE ATTEMPTING TO CLEAR A REACTION TUBE JAM ON THE ANALYZER. THE SAMPLE PROBE WENT INTO THE INDEX FINGER OF ONE HAND. THE OPERATOR OPENED THE LIGHT SHIELD LID TO CLEAR A REACTION TUBE JAM. THE TECHNOLOGIST STATED THAT BECAUSE THE LIGHT SHIELD LID HAD BEEN OPENED HE DID NOT EXPECT THE ANALYZER TO MOVE, BUT THERE WAS MECHANICAL MOVEMENT WITH THE LID OPEN. THIS IS WHEN THE INJURY OCCURRED. THE OPERATOR WENT TO THE EMERGENCY ROOM AFTER THE INCIDENT AT 4:00PM AND WAS TREATED WITH A TETANUS SHOT AND BLOOD TESTS FOR (B)(6). IT IS UNKNOWN IF ANTI-VIRAL MEDICATIONS WERE PRESCRIBED. THE INJURED TECHNOLOGIST WAS TOLD NOT TO WORK FOR A FEW DAYS AFTER THE INJURY BUT, WENT BACK TO WORK AND FINISHED HIS REMAINING SHIFT. A CUSTOMER SERVICE ENGINEER (CSE) WAS DISPATCHED AND ARRIVED AT FACILITY ON 2/25/16 AT 10:00AM. THE ENGINEER REPLACED THE SAMPLE PROBE; REMOVED THE VESSEL FEEDER AND TIGHTENED THE SET SCREW ON THE MOTOR SHAFT. THE ENGINEER EXAMINED THE LIGHT SHIELD LID SENSOR ON THIS ANALYZER ((B)(4)) AND THE LAB'S ALTERNATE CA-1500 ANALYZER ((B)(4)). THE SENSORS FOR THE LIGHT SHIELD LIDS ON BOTH ANALYZERS WERE WORKING HOWEVER THE METAL TABS THAT MOVE IN AND OUT OF THE SENSOR WERE NOT CLEARING THE SENSOR BEAM. SO WHEN THE LID WAS OPENED, THE ANALYZERS DID NOT REGISTER THE LID AS OPEN, AND STOP MECHANICAL MOVEMENT. THE ENGINEER BENT THE METAL TABS OUTWARD TO RESOLVE THE ISSUE SO THAT WHEN THE LID WAS OPENED, THE TABS WOULD CLEAR THE SENSOR BEAM. HE DOES NOT KNOW HOW THE METAL TABS MAY HAVE BECOME OUT OF ALIGNMENT. THE LIGHT SHIELD LID SENSORS WERE NOT OVERRIDDEN OR INACTIVATED WHEN THE CSE EXAMINED THE SENSORS. THERE IS NO KNOWN HISTORY OF LID SENSOR ISSUES AT THIS ACCOUNT. THE CSE REPORTED THAT THE LID STATUS INDICATOR ON THE MAIN SCREEN WAS RED (NOT GREEN) WHEN THE OPERATOR OPENED THE LID, INDICATING THAT THE LID SHOULD NOT BE OPENED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
169719 SYSMEX CA-1500 CP AUTOMATED BLOOD COAGULATION ANALYZER GKP SYSMEX CORPORATION CA-1500 CP

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention