FDA Adverse Event Death Summary report: N

RESOLVE HALO VEST SHORT LAMSWOOL

MDR report key: 5513714 · Received March 21, 2016

Report

Report Number
1836248-2016-00002
Event Type
Death
Date Received
March 21, 2016
Report Date
March 21, 2016
Manufacturer
OSSUR AMERICAS, INC.
Product Code
KQZ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE VEST CURRENTLY PROVIDES INSTRUCTIONS FOR EMERGENCY ACCESS TO THE CHEST AS SHOWN IN THE PHOTO ATTACHED. FURTHER DETAILS WILL BE GATHERED FROM THE HOSPITAL AND WE WILL PROVIDE A FOLLOW UP REPORT.

Additional Manufacturer Narrative · 1

THE VEST CURRENTLY PROVIDES INSTRUCTIONS FOR EMERGENCY ACCESS TO THE CHEST AS SHOWN IN THE PHOTO ATTACHED. WE HAVE INITIATED A PREVENTATIVE ACTION TO MAXIMIZE THE VISIBILITY OF THE EMERGENCY LABEL AND/OR CHANGE THE LABELS TO SYMBOLS RATHER THAN TEXT. -

Description of Event or Problem · 1

PATIENT WAS WEARING HALO VEST AND REQUIRED EMERGENCY CPR. THE STAFF DID NOT KNOW THE CORRECT PROCEDURE TO REMOVE THE VEST TO ADMINISTER CPR. THE PATIENT DIED.

Description of Event or Problem · 1

PATIENT WAS WEARING HALO VEST AND REQUIRED EMERGENCY CPR. THE STAFF DID NOT KNOW THE CORRECT PROCEDURE TO REMOVE THE VEST TO ADMINISTER CPR. THE PATIENT DIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
169693 RESOLVE HALO VEST SHORT LAMSWOOL COMPONENT, TRACTION, NON-INVASIVE KQZ OSSUR AMERICAS, INC. 545700L

Patients

Seq Age Sex Outcome Treatment
1 Death