FDA Adverse Event Malfunction Summary report: N

PERFECTO2 V WITH SENSOR 9153650799

MDR report key: 5513583 · Received March 21, 2016

Report

Report Number
1031452-2016-01016
Event Type
Malfunction
Date Received
March 21, 2016
Report Date
March 3, 2016
Manufacturer
INVACARE FLORIDA OPERATIONS
Product Code
CAW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

JP (B)(6) 2016 THIS PRODUCT WAS EVALUATED AND REPAIRED BY AN INDEPENDENT REPAIR CENTER.  SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL RECORD WILL BE FILED.

Description of Event or Problem · 1

PER THE INDEPENDENT REPAIR CENTER, THE CUSTOMER ALLEGED PROBLEM IS THE UNIT ALARMS NOT FUNCTIONAL. THE KEY FAILURE IS THE POWER SWITCH HAS NO ALARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
171118 PERFECTO2 V WITH SENSOR 9153650799 GENERATOR, OXYGEN, PORTABLE CAW INVACARE FLORIDA OPERATIONS IRC5PO2V

Patients

Seq Age Sex Outcome Treatment
1