FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 300P AND PAD PAK

MDR report key: 5510106 · Received March 18, 2016

Report

Report Number
3004123209-2016-00308
Event Type
Malfunction
Date Received
March 18, 2016
Date of Event
March 7, 2016
Report Date
May 12, 2016
Manufacturer
HEARTSINE TECHNOLOGIES LTD
Product Code
MKJ
PMA / PMN Number
K041067
Removal / Correction Number
Z-0124-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2015058 HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER). H3 OTHER TEXT : NOT YET RETURNED TO MANUFACTURER

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2015058. HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER). THE HISTORY LOG FOR THIS DEVICE SHOWED THAT THE PAD-PAK WAS FIRST INSTALLED ON THE 18TH NOVEMBER 2010 AND THAT IT HAD PERFORMED ALL WEEKLY AUTO SELF-TESTS, ALONGSIDE TWO MANUAL POWER ONS, UP TO THE 19TH JULY 2015. DURING THAT PERIOD THE DEVICE RECORDS MULTIPLE OCCASIONS IN WHICH THE TEMPERATURE WAS RECORDED BELOW THE MINIMUM RECOMMENDED STAND-BY TEMPERATURE OF 0 DEGREES CELSIUS WITH A LOW OF -5 DEGREES CELSIUS RECORDED. MANUAL POWER UPS OF TEN MINUTES WERE RECORDED BETWEEN THE 21ST JULY 2015 AND THE LAST LOG ENTRY ON THE 8TH MARCH 2016. INVESTIGATION FOUND THE REPORTED FAULT CAN BE ATTRIBUTED TO MEMBRANE FAILURE. THE TEN MINUTE TIME OUTS AND THE MEASUREMENTS TAKEN WOULD SUGGEST A FAILING MEMBRANE. THE FAULT COULD NOT BE REPLICATED WITH A NEW MEMBRANE FITTED. THE PAD-PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.

Description of Event or Problem · 0

THERE WAS NO PATIENT INVOLVED IN THIS EVENT. UNIT POWERS ON BY ITSELF WITHOUT MANUAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
167789 HEARTSINE SAMARITAN 300P AND PAD PAK AUTOMATED EXTERNAL DEFIBRILLATOR MKJ HEARTSINE TECHNOLOGIES LTD SAM N/A

Patients

Seq Age Sex Outcome Treatment
1