FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH ULTRA
MDR report key: 550970
·
Received April 12, 2004
Report
- Report Number
- 2939301-2004-02348
- Event Type
- Malfunction
- Date Received
- April 12, 2004
- Report Date
- April 8, 2004
- Manufacturer
- LIFESCAN, INC.
- Product Code
- CFR
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
Description of Event or Problem · 1
A PT REPORTED BLOOD GLUCOSE RESULTS OF 20.9 MMO/ AND 12.1 MMO/L WITH A LIFESCAN METER, PERFORMED WITHIN 10 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS THE EXPECTED VALUE OF <= 20% AND/OR 20 MG/DL, THEREBY INDICATING A POTENTIAL MALFUNCTION OF THE METER IN TERMS OF PRECISION. A CONTROL SOLUTION TEST WAS PERFORMED AND THE RESULT FELL WITHIN SPECIFICATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH ULTRA | GLUCOSE MONITORING SYS/KI | CFR | LIFESCAN, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR |