IPC® HANDPIECE - INDIGO¿ DRILL
Report
- Report Number
- 1045254-2016-00076
- Event Type
- Malfunction
- Date Received
- March 17, 2016
- Report Date
- February 23, 2016
- Manufacturer
- MEDTRONIC XOMED INC.
- Product Code
- HBE
- PMA / PMN Number
- K081475
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
THIS DEVICE IS USED FOR THERAPEUTIC PURPOSES. CONCOMITANT MEDICAL PRODUCTS: INDIGO DRILL #1845000, SN# (B)(4), LOT# 210240093; INDIGO DRILL #1845000, SN# (B)(4), LOT# 210150121; INDIGO ATTACHMENT #1845020, SN# (B)(4), LOT# 210302642; INDIGO ATTACHMENT #1845020, SN# (B)(4), LOT# 208777604; INDIGO ATTACHMENT #1845020, SN# (B)(4), LOT# 210302644. (B)(4). THE DEVICES HAVE NOT BEEN RECEIVED FOR EVALUATION.
DEVICE AVAILABLE FOR EVAL: 03/24/2016. THE DEVICES WERE RETURNED AND ANALYZED AS FOLLOWS: INDIGO DRILL #1845000 SN# (B)(4) LOT# 208868102: ANALYSIS FOUND NO FUNCTIONAL FAULT WITH THE DEVICE AND THE REPORTED OVERHEATING WAS NOT CONFIRMED. TEMPERATURE TEST AT ONE MINUTE WAS 76.1°F. VISUAL ANALYSIS FOUND THE DEVICE IS DISCOLORED. THE DRILL WAS CLEANED AND TESTED TO PRODUCT SPECIFICATIONS AND RETURNED TO CUSTOMER. INDIGO DRILL #1845000 SN# (B)(4) LOT# 208796687: ANALYSIS FOUND NO FUNCTIONAL FAULT WITH THE DEVICE AND THE REPORTED OVERHEATING WAS NOT CONFIRMED. TEMPERATURE TEST AT ONE MINUTE WAS 76°F. VISUAL ANALYSIS FOUND THE DEVICE IS DISCOLORED. THE DRILL WAS CLEANED AND TESTED TO PRODUCT SPECIFICATIONS AND RETURNED TO CUSTOMER. INDIGO DRILL #1845000 SN# (B)(4) LOT# 209592797: ANALYSIS FOUND NO FUNCTIONAL FAULT WITH THE DEVICE AND THE REPORTED OVERHEATING WAS NOT CONFIRMED. TEMPERATURE TEST AT ONE MINUTE WAS 75.7°F. VISUAL ANALYSIS FOUND THE DEVICE IS DISCOLORED. THE DRILL WAS CLEANED AND TESTED TO PRODUCT SPECIFICATIONS AND RETURNED TO CUSTOMER. INDIGO ATTACHMENT #1845020 SN# (B)(4) LOT# 209752490: ANALYSIS FOUND NO FUNCTIONAL FAULT WITH THE DEVICE AND THE REPORTED OVERHEATING WAS NOT CONFIRMED. TEMPERATURE TEST AT ONE MINUTE WAS 74.1°F. THE DEVICE WAS CLEANED AND TESTED TO PRODUCT SPECIFICATIONS AND RETURNED TO CUSTOMER. INDIGO ATTACHMENT #1845020 SN# (B)(4) LOT# 208777604: ANALYSIS FOUND NO FUNCTIONAL FAULT WITH THE DEVICE AND THE REPORTED OVERHEATING WAS NOT CONFIRMED. TEMPERATURE TEST AT ONE MINUTE WAS 75.8°F. THE DEVICE WAS CLEANED AND TESTED TO PRODUCT SPECIFICATIONS AND RETURNED TO CUSTOMER. INDIGO ATTACHMENT #1845020 SN# (B)(4) LOT# 209797053: ANALYSIS FOUND NO FUNCTIONAL FAULT WITH THE DEVICE AND THE REPORTED OVERHEATING WAS NOT CONFIRMED. TEMPERATURE TEST AT ONE MINUTE WAS 76.8°F. THE DEVICE WAS CLEANED AND TESTED TO PRODUCT SPECIFICATIONS AND RETURNED TO CUSTOMER. (B)(4).
CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE DRILLS AND ATTACHMENTS WERE NOT WORKING PROPERLY AND HEATING UP. IT WAS NOTED THAT NONE OF THE ISSUES OCCURRED DURING A SURGERY. THERE WAS NO REPORT OF INJURY AS A RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 164107 | IPC® HANDPIECE - INDIGO¿ DRILL | DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) | HBE | MEDTRONIC XOMED INC. | 1845000 | 208868102 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |