HILL-ROM AIRSHIELDS ISOLETTE INFANT INCUBATOR

MDR report key: 550756 · Received August 31, 2004

Report

Report Number
550756
Event Type
Other
Date Received
August 31, 2004
Date of Event
August 26, 2004
Report Date
August 31, 2004
Manufacturer
HILL-ROM MANUFACTURING, INC.
Product Code
FMZ
Product Problem
Yes
Report Source
User Facility report
Reporter Location
PA, US
Reporter Occupation
INVALID DATA

Narratives

Description of Event or Problem · 1

A PATIENT ROLLED AGAINST AN UNSECURED PORT DOOR ON AN AIR SHIELDS ISOLETTE (MODEL C100) AND FELL OUT OF THE ISOLETTE. AS THE PATIENT ROLLED OUT OF THE PORT DOOR, THE PULSE OXIMETER'S SENSOR ATTACHED TO THE PATIENT'S FOOT PREVENTED, THE PATIENT FROM REACHING THE FLOOR. THE RESULTS OF AN IMMEDIATE EXAMINATION OF THE PATIENT CONCLUDED THAT THERE WERE NO INJURIES AS A RESULT OF THE FALL. THE ISOLETTE WAS REMOVED FROM SERVICE. BIOMEDICAL ENGINEERING EXAMINED THE ISOLETTE AND IDENTIFIED THAT THE LATCH ON THE OLDER STYLE C100 PORT DOORS HAVE TO BE PUSHED UNTIL A CLICK IS HEARD WHICH INDICATES THAT THE DOOR IS SECURELY CLOSED. THE DOOR CAN BE HELD IN PLACE WITHOUT BEING FULLY SECURED WHICH WAS THE CAUSE OF THE FALL. BIOMEDICAL ENGINEERING IDENTIFIED THREE OTHER C100 ISOLETTES WITH THIS TYPE OF LATCH AND REMOVED THEM FROM SERVICE. BIOMEDICAL ENGINEERING HAS REPLACED THE LATCHES ON TWO OF THE C100 ISOLETTES WITH NEWER STYLE LATCHES. THE NEWER STYLE LATCHES REQUIRE LESS EFFORT TO CLOSE AND SECURE THE DOOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HILL-ROM AIRSHIELDS ISOLETTE INFANT INCUBATOR ISOLETTE INFANT INCUBATOR FMZ HILL-ROM MANUFACTURING, INC. C100/200-2 *

Patients

Seq Age Sex Outcome Treatment
1 *