FDA Adverse Event Malfunction Summary report: N

EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE

MDR report key: 5506114 · Received March 16, 2016

Report

Report Number
2951238-2016-00313
Event Type
Malfunction
Date Received
March 16, 2016
Date of Event
February 19, 2016
Report Date
March 1, 2019
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
ODG
PMA / PMN Number
PK093395
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE REFERENCED IN THIS REPORT HAS NOT BEEN RETURNED TO OLYMPUS FOR EVALUATION. AS PART OF OUR INVESTIGATION INTO THIS REPORT, OLYMPUS DISPATCHED AN ENDOSCOPY SUPPORT SPECIALIST (ESS) TO THE FACILITY TO OBSERVE THEIR REPROCESSING PRACTICES. ESS VISITED THE FACILITY ON 2/22/2016 TO REVIEW THE REPROCESSING STEPS OF THE GF-UCT180 BASED ON THE REPROCESSING MANUAL. IT WAS IDENTIFIED THAT SCOPE SUCTIONING PROCESS DURING PRE-CLEANING WAS NOT BEING PERFORMED, AND THE PROPER BRUSH FOR CLEANING WAS NOT BEING UTILIZED. ESS RECOMMENDED BOTH STEPS BE IMPLEMENTED AND TO SCHEDULE AN ADDITIONAL IN-SERVICE VISIT IN THE FUTURE. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED. HOWEVER, INSUFFICIENT REPROCESSING COULD BE A CONTRIBUTING FACTOR.

Additional Manufacturer Narrative · 1

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO MAKE A CORRECTION ON THE PROCODE FROM NWB TO ODG.

Description of Event or Problem · 1

OLYMPUS WAS INFORMED THAT AFTER REPROCESSING THE SCOPE IN A CUSTOM ULTRASONICS AUTOMATED ENDOSCOPE REPROCESSOR (AER), TRACES OF BIO-BURDEN MATERIAL WAS NOTED AT THE DISTAL TIP OF THE DEVICE. THE FACILITY REPORTED THAT THE SCOPE FAILED A CULTURE TEST FOR AN UNSPECIFIED BACTERIA. THE FACILITY PLANS TO HAVE THE SCOPE FURTHER REPROCESSED USING A MEDIVATORS AER. THE SCOPE WILL NOT BE RETURNED FOR FURTHER EVALUATION. THERE'S NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
161558 EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE ULTRASOUND GASTROVIDEOSCOPE ODG OLYMPUS MEDICAL SYSTEMS CORP. GF-UCT180

Patients

Seq Age Sex Outcome Treatment
1