FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 5500875
·
Received March 15, 2016
Report
- Report Number
- 3008642652-2016-01490
- Event Type
- Malfunction
- Date Received
- March 15, 2016
- Date of Event
- February 20, 2016
- Report Date
- March 11, 2016
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (NON-FUNCTIONAL THERAPY ELECTRODES) HAS BEEN CONFIRMED. UPON INVESTIGATION, THE ELECTRODE BELT FAILED A THERAPY ELECTRODE RECOGNITION TEST. THE CAUSE FOR THE FAILURE WAS ISOLATED TO AN OPEN PULSE WIRE IN THE CABLE CONNECTING ECG "A" AND ECB "B." THE ROOT CAUSE FOR THE OPEN WIRE WAS EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT.
Description of Event or Problem · 1
A US DISTRIBUTOR RETURNED ELECTRODE BELT SN (B)(4) AND REPORTED THAT THE ELECTRODE BELT HAD NON-FUNCTIONAL THERAPY ELECTRODES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 156755 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | WCD 4000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |