FDA Adverse Event Death Summary report: N

PANORAMA CENTRAL STATION WITH TELEMETRY

MDR report key: 5500581 · Received March 15, 2016

Report

Report Number
3007222337-2016-00007
Event Type
Death
Date Received
March 15, 2016
Date of Event
February 9, 2016
Report Date
March 15, 2016
Manufacturer
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO. LTD
Product Code
MHX
PMA / PMN Number
K150082
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

COMPANY REPRESENTATIVE EVALUATED THE SYSTEM AND REPORTS THE SYSTEM DID NOT HAVE AUDIBLE ALARM DUE TO A CABLE THAT WAS NOT COMPLETELY PLUGGED INTO THE DEVICE. THERE WERE NO DEVICE RELATED MALFUNCTION. MINDRAY DS USA, INC. ((B)(4)) IS SUBMITTING THIS REPORT ON BEHALF OF (B)(4) MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. ((B)(4)). (EXEMPTION #E2015032).

Description of Event or Problem · 0

CUSTOMER REPORTED AN ISSUE WITH THE PANORAMA CENTRAL STATION WITH TELEMETRY WHERE A PATIENT EXPIRED AND THERE WERE NO AUDIBLE ALARMS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
156895 PANORAMA CENTRAL STATION WITH TELEMETRY CENTRAL MONITORING STATION MHX SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO. LTD

Patients

Seq Age Sex Outcome Treatment
1 Death