SILVERHAWK BTK
Report
- Report Number
- 2183870-2016-00249
- Event Type
- Malfunction
- Date Received
- March 14, 2016
- Date of Event
- May 12, 2015
- Report Date
- May 14, 2015
- Manufacturer
- COVIDIEN
- Product Code
- MCW
- PMA / PMN Number
- K053460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS MDR IS BEING SUBMITTED AS A PART OF A RETROSPECTIVE REVIEW / REMEDIATION EFFORT PERFORMED AT THE COVIDIEN (B)(4) LOCATION, FOLLOWING MEDTRONIC¿S ACQUISITION OF COVIDIEN. A CAPA HAS BEEN OPENED TO MANAGE THE ACTIONS RELATED TO REMEDIATION OF COMPLAINT FILES AND ANY REQUIRED MDR REPORTING. (B)(4).
IT IS REPORTED THAT THE DEVICE WOULD NOT PASS WHEN PUSHED THROUGH THE LESION AND WHEN THE DEVICE WAS BROUGHT OUT AND THE LESION WAS PRE DILATED IT WAS NOTICED THAT THE DISTAL TIP WAS SPLIT. THE PHYSICIAN ALSO NOTED A KINK NEAR THE HUB SO REMOVED THE DEVICE AND COMPLETED THE PROCEDURE WITH A NEW DEVICE. NO PATIENT INJURY IS REPORTED. EVALUATION SUMMARY: THE SILVERHAWK DEVICE WAS RECEIVED FOR EVALUATION WITHOUT THE CUTTER DRIVER (ATTACHED) OR ANY OTHER ANCILLARY DEVICES. THE DISTAL TIP ASSEMBLY HOUSING EXHIBITED ASYMMETRICAL DEFORMATION THAT COULD HAVE BEEN EITHER PRESENT DURING THE PROCEDURE OR GENERATED POST-PROCEDURE. THE POLYIMIDE GUIDEWIRE LUMEN (GWL) WAS FRACTURED AT THE DISTAL EDGE OF THE TIP ASSEMBLY HOUSING. THE GWL PROXIMAL TO THE FRACTURE SITE EXHIBITED A LONGITUDINAL TEAR FROM THE PROXIMAL EDGE TO THE FRACTURE. THE GWL DISTAL TO THE FRACTURE HAD PEELED FROM THE TECOTHANE JACKET UNTIL THE GWL WAS INCORPORATED DEEP ENOUGH INTO THE TECOTHANE TAPERED TIP TO STOP THE DAMAGE. THE WHITE TORQUE SHAFT SLEEVE WAS FULLY INTACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 155606 | SILVERHAWK BTK | CATHETER, PERIPHERAL, ATHERECTOMY | MCW | COVIDIEN | P4028 | A064593 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |