FDA Adverse Event Malfunction Summary report: N

SILVERHAWK BTK

MDR report key: 5498592 · Received March 14, 2016

Report

Report Number
2183870-2016-00249
Event Type
Malfunction
Date Received
March 14, 2016
Date of Event
May 12, 2015
Report Date
May 14, 2015
Manufacturer
COVIDIEN
Product Code
MCW
PMA / PMN Number
K053460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS MDR IS BEING SUBMITTED AS A PART OF A RETROSPECTIVE REVIEW / REMEDIATION EFFORT PERFORMED AT THE COVIDIEN (B)(4) LOCATION, FOLLOWING MEDTRONIC¿S ACQUISITION OF COVIDIEN. A CAPA HAS BEEN OPENED TO MANAGE THE ACTIONS RELATED TO REMEDIATION OF COMPLAINT FILES AND ANY REQUIRED MDR REPORTING. (B)(4).

Description of Event or Problem · 1

IT IS REPORTED THAT THE DEVICE WOULD NOT PASS WHEN PUSHED THROUGH THE LESION AND WHEN THE DEVICE WAS BROUGHT OUT AND THE LESION WAS PRE DILATED IT WAS NOTICED THAT THE DISTAL TIP WAS SPLIT. THE PHYSICIAN ALSO NOTED A KINK NEAR THE HUB SO REMOVED THE DEVICE AND COMPLETED THE PROCEDURE WITH A NEW DEVICE. NO PATIENT INJURY IS REPORTED. EVALUATION SUMMARY: THE SILVERHAWK DEVICE WAS RECEIVED FOR EVALUATION WITHOUT THE CUTTER DRIVER (ATTACHED) OR ANY OTHER ANCILLARY DEVICES. THE DISTAL TIP ASSEMBLY HOUSING EXHIBITED ASYMMETRICAL DEFORMATION THAT COULD HAVE BEEN EITHER PRESENT DURING THE PROCEDURE OR GENERATED POST-PROCEDURE. THE POLYIMIDE GUIDEWIRE LUMEN (GWL) WAS FRACTURED AT THE DISTAL EDGE OF THE TIP ASSEMBLY HOUSING. THE GWL PROXIMAL TO THE FRACTURE SITE EXHIBITED A LONGITUDINAL TEAR FROM THE PROXIMAL EDGE TO THE FRACTURE. THE GWL DISTAL TO THE FRACTURE HAD PEELED FROM THE TECOTHANE JACKET UNTIL THE GWL WAS INCORPORATED DEEP ENOUGH INTO THE TECOTHANE TAPERED TIP TO STOP THE DAMAGE. THE WHITE TORQUE SHAFT SLEEVE WAS FULLY INTACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
155606 SILVERHAWK BTK CATHETER, PERIPHERAL, ATHERECTOMY MCW COVIDIEN P4028 A064593

Patients

Seq Age Sex Outcome Treatment
1