FDA Adverse Event Injury Summary report: N

MICRO TOUCH PLUS PF

MDR report key: 5494494 · Received March 11, 2016

Report

Report Number
1019632-2016-00002
Event Type
Injury
Date Received
March 11, 2016
Date of Event
February 22, 2016
Report Date
June 22, 2016
Manufacturer
SIAM SEMPERMED CORP LTD
Product Code
LYY
Adverse Event
Yes
Report Source
Distributor report
Reporter Location
NY, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4) CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD. THE REVIEW INDICATED THAT THE LOT WAS IN COMPLIANCE WITH RELEASE CRITERIA. RETURNED SAMPLES WERE TESTED FOR TOTAL PROTEIN CONTENT, PH, AND POWDER RESIDUE WITH RESULTS WITHIN SPECIFICATION. FROM THE TESTING, THERE WAS NO INDICATION FOR A CAUSE OF THE CITED REACTION. PART OF THE RETURNED SAMPLES WERE SENT TO AN EXTERNAL TESTING LAB FOR ADDITIONAL TESTING. UPON COMPLETION, RESULTS WILL BE FORWARDED BACK TO ANSELL. ON 06/22/2016 UPDATE: AN EXTERNAL LAB ANALYSIS OF THE GLOVE CONFIRMED THAT THE LEVELS OF ANTI OXIDANTS AND ADDITIVES ON THE RETURNED SAMPLES WERE WITHIN SPECIFICATION IN COMPARISON TO HISTORICAL DATA ((B)(4)).

Additional Manufacturer Narrative · 1

SIMA SEMPERMED CORP., LTD., LOCATED IN (B)(4) CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD. THE REVIEW INDICATED THAT THE LOT WAS IN COMPLIANCE WITH RELEASE CRITERIA. RETURNED SAMPLES WERE TESTED FOR TOTAL PROTEIN CONTENT, PH, AND POWDER RESIDUE WITH RESULTS WITHIN SPECIFICATION. FROM THE TESTING, THERE WAS NO INDICATION FOR A CAUSE OF THE CITED REACTION. PART OF THE RETURNED SAMPLES WERE SENT TO AN EXTERNAL TESTING LAB FOR ADDITIONAL TESTING. UPON COMPLETION, RESULTS WILL BE FORWARDED BACK TO (B)(4).

Description of Event or Problem · 1

CUSTOMER INDICATED THAT BACK OF HER HANDS WERE BLEEDING AFTER WEARING THESE GLOVES. SHE HAS WORN THESE TYPE OF GLOVES FOR 35 YEARS AND NEVER HAD A PROBLEM. ONCE SHE SWITCHED TO NITRILE GLOVES HER HANDS GOT BETTER ON 06/22/2016 UPDATE: AN EXTERNAL LAB ANALYSIS OF THE GLOVE CONFIRMED THAT THE LEVELS OF ANTI OXIDANTS AND ADDITIVES ON THE RETURNED SAMPLES WERE WITHIN SPECIFICATION IN COMPARISON TO HISTORICAL DATA ((B)(4)).

Description of Event or Problem · 1

CUSTOMER INDICATED THAT BACK OF HER HANDS WERE BLEEDING AFTER WEARING THESE GLOVES. SHE HAS WORN THESE TYPE OF GLOVES FOR 35 YEARS AND NEVER HAD A PROBLEM. ONCE SHE SWITCHED TO NITRILE GLOVES HER HANDS GOT BETTER.

Description of Event or Problem · 1

CUSTOMER INDICATED THAT BACK OF HER HANDS WERE BLEEDING AFTER WEARING THESE GLOVES. SHE HAS WORN THESE TYPE OF GLOVES FOR 35 YEARS AND NEVER HAD A PROBLEM. ONCE SHE SWITCHED TO NITRILE GLOVES HER HANDS GOT BETTER

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
151579 MICRO TOUCH PLUS PF MICRO TOUCH PLUS PF LATEX EXAMINATION GLOVE LYY SIAM SEMPERMED CORP LTD STYLE 530 1503000977

Patients

Seq Age Sex Outcome Treatment
1 55 YR Other