PLMA DVC V11.51 1 N CR
Report
- Report Number
- 9615050-2016-00037
- Event Type
- Malfunction
- Date Received
- March 10, 2016
- Date of Event
- March 3, 2016
- Report Date
- March 4, 2016
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K070398
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED.
TESTING AND INVESTIGATION FOUND DURING VISUAL INSPECTION THAT THE DISTAL PRESSURE PIN ON THE PRESSURE DETECTOR ASSEMBLY WAS BROKEN. THE DEVICE DID NOT PASS THE DISTAL OCCLUSION TEST; HOWEVER THE DEVICE AUDIBLY ALARMED FOR A DISTAL OCCLUSION DURING THE 6 PSI SETPOINT AT 25.63 PSI. THE PROBABLE CAUSE WAS DUE TO PHYSICAL DAMAGE TO THE PRESSURE DETECTOR ASSEMBLY.
THE CUSTOMER CONTACT REPORTED THAT DURING TESTING AT THE USER FACILITY, THE DEVICE DID NOT PASS THE DISTAL OCCLUSION TEST BY CONTINUING TO INFUSE PAST 25 PSI AT THE 6 PSI SETPOINT. IT WAS FURTHER REPORTED THAT THE CUSTOMER CONFIRMED THE DEVICE DID NOT DETECT A DISTAL OCCLUSION. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFORMATION WAS PROVIDED.
THE CUSTOMER CONTACT REPORTED THAT DURING TESTING AT THE USER FACILITY, THE DEVICE DID NOT PASS THE DISTAL OCCLUSION TEST BY CONTINUING TO INFUSE PAST 25 PSI AT THE 6 PSI SETPOINT. IT WAS FURTHER REPORTED THAT THE CUSTOMER CONFIRMED THE DEVICE DID NOT DETECT A DISTAL OCCLUSION. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 147745 | PLMA DVC V11.51 1 N CR | PUMP, INFUSION | FRN | HOSPIRA COSTA RICA LTD. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |