PRODIGY AUTOCODE
Report
- Report Number
- 3005862821-2016-00010
- Event Type
- Injury
- Date Received
- March 10, 2016
- Date of Event
- January 23, 2016
- Report Date
- January 29, 2016
- Manufacturer
- OK BIOTECH CO., LTD.
- Product Code
- NBW
- PMA / PMN Number
- K073118
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
SUSPECTED DEVICE AND RETURNED STRIPS EVALUATED BY OK BIOTECH AND CALCULATED THAT THE METER OPERATED WITHIN SPECIFICATIONS AND RETURNED STRIPS ARE AT THE SAME PERFORMANCE WITH OUR RETAINED STRIPS. ON DEVICE TESTING: THE STANDBY CURRENT TEST IS 1.5A. THE CRITERIA IS <55UA, PASS. METER SETTING, AUDIO AND ALL BUTTONS FUNCTION ARE OK. WE TESTED THE SUSPECTED METER WITH IN HOUSE CONTROL SOLUTION AND RETAINED STRIPS, SAME STRIP LOT D150806-2 THAT PATIENT RETURNED. THE CONTROL SOLUTION TESTS FOR LEVEL LOW ARE 60/61 MG/DL, FOR LEVEL HIGH ARE 281/278 MG/DL. THE CONTROL SOLUTION RANGES ARE: LEVEL LOW 25~70 MG/DL; LEVEL HIGH 210~310 MG/DL. ALL RESULTS WERE WITHIN THE ACCEPTANCE RANGE. PASS. WE ALSO TESTED THE RETURNED STRIPS FROM PATIENT (STRIP LOT NUMBER: D150806-2). THE CONTROL SOLUTION TESTS FOR LEVEL LOW WERE 66/64 MG/DL; FOR LEVEL HIGH WERE 276/271 MG/DL.ALL RESULTS WERE WITHIN THE ACCEPTANCE RANGE. PASS THE PERFORMANCE OF OUR RETAINED STRIPS AND PATIENT RETURNED STRIPS ARE ALL THE SAME. HENCE FIND NO FAILURE ON BOTH METER AND STRIPS. (B)(4).
OUR IMPORTER, PRODIGY DIABETES CARE, RECEIVED A CALL ON (B)(6) 2016 REPORTING A MEDICAL INTERVENTION THAT OCCURRED ON (B)(6) 2016 AT 11PM. PATIENT'S WIFE ((B)(6)) CALLED IN STATING THAT PATIENT BLOOD GLUCOSE TESTED HIGHER THAN NORMAL AND WAS UNABLE TO GET UP FROM A NAP. PATIENT WAS EXPERIENCING SYMPTOMS OF FATIGUE AND COULD NOT SPEAK CLEARLY. THE READING ON THE PRODIGY METER AT THE TIME OF THE EVENT WAS 528MG/DL. PATIENT'S NORMAL BLOOD GLUCOSE READING AT THE TIME OF DAY OF THE EVENT IS 130MG/DL. PARAMEDICS WERE CALLED 30 MINUTES AFTER TESTING WITH THE PRODIGY METER. PARAMEDICS ARRIVED 10-15 MINUTES AFTER BEING CALLED. APPROXIMATELY 30-45 MINUTES PASSED BETWEEN TESTING WITH THE PRODIGY METER AND THE PARAMEDIC'S METER. PARAMEDICS TESTED PATIENT'S GLUCOSE WITH THEIR METER BUT ONLY STATED THAT THE RESULTS WERE LOW. PATIENT WAS TRANSPORTED TO ER BY PARAMEDICS. UPON ARRIVAL AT ER PATIENT'S BLOOD GLUCOSE WAS NOT REGISTERING BECAUSE IT WAS TOO LOW. PATIENT'S GLUCOSE UPON DISCHARGE WAS 130MG/DL. IT WAS UNKNOWN WHAT TREATMENT PATIENT RECEIVED AT ER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 148996 | PRODIGY AUTOCODE | BLOOD GLUCOSE MONITORING DEVICE | NBW | OK BIOTECH CO., LTD. | 51850 | D150806-2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | Required Intervention | FLOMAX 0.4MG |