FDA Adverse Event Injury Summary report: N

THERMOCOOL® SMARTTOUCH® SF UNI-DIRECTIONAL NAV CATHETER

MDR report key: 5488176 · Received March 8, 2016

Report

Report Number
9673241-2016-00177
Event Type
Injury
Date Received
March 8, 2016
Date of Event
November 30, 2015
Report Date
December 1, 2015
Manufacturer
BIOSENSE WEBSTER, INC. (JUAREZ)
Product Code
ILO
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ¿SUSPECTED MEDICAL DEVICE¿ REPORTED IN THIS REPORT IS NOT MARKETED IN USA OR APPROVED BY THE FDA. HOWEVER, IT IS BEING REPORTED AS BIOSENSE WEBSTER CONSIDERS THIS AS A SIMILAR DEVICE TO THERMOCOOL SMART TOUCH BI-DIRECTIONAL NAVIGATION CATHETER APPROVED UNDER P030031/S053. (B)(4) ARE RELATED TO THE SAME INCIDENT. UPON RECEIPT, THE CATHETER WAS VISUALLY INSPECTED AND IT WAS FOUND IN NORMAL CONDITIONS. THE RETURNED DEVICE WAS THEN EVALUATED FOR ELECTRICAL RESISTANCE AND THE THERMOCOUPLE TEST FAILED. FURTHER EXAMINATION REVEALED THAT THE THERMOCOUPLE WIRES WERE BROKEN AT THE TIP SECTION CREATING AN INTERMITTENCE OF TEMPERATURE. AN INTERNAL CORRECTIVE ACTION WAS CREATED TO ADDRESS THIS ISSUE. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. THE CATHETER FAILED DURING THERMOCOUPLE TEST HOWEVER THIS CONDITION IS NOT RELATED TO THE THROMBOSIS AND PULMONARY VEIN STENOSIS. THE ROOT CAUSE OF THE THERMOCOUPLE BREAKAGE COULD NOT BE DETERMINED.

Additional Manufacturer Narrative · 1

ADDITIONAL TESTING WAS PERFORMED ON THE CATHETER AFTER THE INITIAL 3500A REPORT WITH THE PRODUCT ANALYSIS WAS SENT. (B)(4) IT WAS REPORTED THAT A (B)(6) MALE PATIENT UNDERWENT AN ABLATION PROCEDURE FOR ATRIAL FIBRILLATION WITH A THERMOCOOL SMARTTOUCH SF UNI-DIRECTIONAL NAV CATHETER AND SUFFERED PULMONARY VEIN STENOSIS AND THROMBOSIS. THE RETURNED DEVICE WAS VISUALLY AND IT WAS FOUND IN NORMAL CONDITIONS. AN IRRIGATION TEST WAS PERFORMED AND THE CATHETER PASSED, NO OCCLUSION WAS OBSERVED. A DEFLECTION TEST WAS PERFORMED AND THE CATHETER PASSED. THE RETURNED DEVICE WAS THEN EVALUATED FOR ELECTRICAL RESISTANCE AND THERMOCOUPLE TEST. CATHETER FAILED DURING THERMOCOUPLE TEST. FURTHER EXAMINATION REVEALED THAT THE THERMOCOUPLE WIRES WERE BROKEN AT THE TIP SECTION. PER THIS CONDITION FORCE FEATURE CANNOT BE VERIFIED; HOWEVER THE CATHETER WAS RECOGNIZED BY CARTO 3 SYSTEM, NO ERROR MESSAGES WERE DISPLAYED AND THE CATHETER WAS PROPERLY VISUALIZED. EEPROM DATA DEMONSTRATES THE CATHETER WAS PROPERLY CALIBRATED DURING MANUFACTURING. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. CATHETER FAILED DURING THERMOCOUPLE TEST; HOWEVER THIS CONDITION IS NOT RELATED TO THE PULMONARY VEIN STENOSIS AND THROMBOSIS LATER REPORTED. THE ROOT CAUSE OF THE OUTCOME PRESENTED ON THE PATIENT REMAINS UNKNOWN. AN INTERNAL CORRECTIVE ACTION WAS CREATED TO ADDRESS TC BREAKAGE ON THE TIP SECTION FOR ST AND ST SF CATHETERS.

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) MALE PATIENT UNDERWENT AN ABLATION PROCEDURE FOR ATRIAL FIBRILLATION WITH A THERMOCOOL SMARTTOUCH SF UNI-DIRECTIONAL NAV CATHETER AND SUFFERED PULMONARY VEIN STENOSIS AND THROMBOSIS. THERE WAS A LOSS OF DISPLAY OF THE TEMPERATURE VALUE. THE CATHETER WAS REPLACED AND THE ISSUE RESOLVED. THE PROCEDURE WAS COMPLETED WITHOUT PATIENT CONSEQUENCE. THIS EVENT WAS ORIGINALLY ASSESSED AS NOT MDR REPORTABLE BECAUSE THE POTENTIAL THAT IT COULD CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY, OR OTHER SIGNIFICANT ADVERSE EVENT, IS REMOTE. POST-PROCEDURE, THE PATIENT RETURNED TO THE HOSPITAL ON (B)(6) 2016 WITH DYSPNEA AND CHEST PAIN ASSOCIATED WITH BREATHING. AN ANGIOGRAPHY REVEALED LEFT INFERIOR PULMONARY VEIN PREOSTIAL STENOSIS AND LEFT SUPERIOR PULMONARY VEIN THROMBOSIS. THE PATIENT'S CONDITION IS NOW BETTER DESPITE AN ADVERSE EVENT RELATED TO ANTICOAGULATION. THERE IS NO FURTHER INFORMATION ABOUT ANY ADDITIONAL INTERVENTION PERFORMED. THE PATIENT LEFT THE HOSPITAL ON (B)(6) 2016. THE PHYSICIAN DID NOT PROVIDE CAUSALITY OPINION FOR THE EVENT. THE AWARENESS DATE FOR THIS RECORD IS FEBRUARY 17, 2016 BECAUSE THAT IS WHEN THE PATIENT UPDATE INFORMATION WAS RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
143142 THERMOCOOL® SMARTTOUCH® SF UNI-DIRECTIONAL NAV CATHETER SIMILAR DEVICE D133601, PMA # P030031/S053 ILO BIOSENSE WEBSTER, INC. (JUAREZ) D-1347-02-S 17175080L

Patients

Seq Age Sex Outcome Treatment
1 52 YR Hospitalization