FDA Adverse Event Malfunction Summary report: N

E SERIES DEFIBRILLATOR

MDR report key: 5486329 · Received March 8, 2016

Report

Report Number
1220908-2016-00515
Event Type
Malfunction
Date Received
March 8, 2016
Report Date
February 17, 2016
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K042007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO AN UNSEATED FLEX CABLE TO THE INTERLOCK CONNECTOR. THE INTERLOCK CONNECTOR AND FLEX CABLE WERE REPLACED TO RESOLVE THE MALFUNCTION. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORPORATION HAS NOT RECEIVED THE DEVICE FOR EVALUATION AND THIS COMPLAINT IS STILL UNDER INVESTIGATION.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "DEFIB FAULT 72" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
143460 E SERIES DEFIBRILLATOR E SERIES MKJ ZOLL MEDICAL CORPORATION E SERIES NA