FDA Adverse Event Injury Summary report: N

CENTRAL BODY F/MEF MAX-L 437 MIN-L 287

MDR report key: 5485681 · Received March 8, 2016

Report

Report Number
3003875359-2016-10166
Event Type
Injury
Date Received
March 8, 2016
Date of Event
January 20, 2016
Report Date
February 8, 2016
Manufacturer
SYNTHES HAGENDORF
Product Code
HST
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

MANUFACTURING INVESTIGATION EVALUATION: ONE (1) ARTICLE OF CENTRAL BODY F/MEF MAX-L 437 MIN-L 287 (PART 392.901) WAS RECEIVED FOR MANUFACTURING INVESTIGATION. THE ARTICLE IS IN A USED CONDITION WITH VISIBLE MARKS OF USE AT THE ANODIZED END PARTS. THE DEPARTMENT RESPONSIBLE FOR THE ASSEMBLY AND TESTING OF THE CENTRAL BODY F/MEF HAS EVALUATED THE COMPLAINED ISSUE. THE INVESTIGATION WAS TO DETERMINE IF THE ISSUE, AS IT WAS REPORTED, COULD BE FOUND WHEN MANIPULATING THE MOVABLE PARTS. THE EXACT MALFUNCTION COULD NOT BE DUPLICATED DURING THE TESTS. HOWEVER, THE FOLLOWING ANOMALIES WERE FOUND: THE SUBCOMPONENT DYNAMIZATION END PIECE (PART 392.901.013) WAS MAKING NOISE AND AN ABNORMAL TYPE OF RESISTANCE WAS FELT WHEN TURNING ON IT. ALSO, DAMAGES ON THE O-RING SEAL, RESIDUES ON THE OUTER THREAD, AND DISCOLORATION OF THE WRENCH SURFACE WERE OBSERVED. IT IS LIKELY THAT THE RESIDUE CAME INTO THE MECHANISM OF THE DYNAMIZATION END PIECE AS A RESULT OF INSUFFICIENT CLEANING AND REPROCESSING. NO MANUFACTURING RELATED ISSUES WERE IDENTIFIED AND/OR CONFIRMED. DEVICE USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. (B)(6). THE EXACT DATE OF IMPLANT IS UNKNOWN AND WAS APPROXIMATELY (B)(6) 2016. (B)(6). A DEVICE HISTORY RECORD REVIEW WAS PERFORMED FOR THE SUBJECT DEVICE LOT. THE REVIEW SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. THE SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. (B)(4).IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE CENTRAL BODY FOR THE MONOLATERAL EXTERNAL FIXATION SYSTEM FOR TRAUMA AND ORTHOPEDICS (MEFISTO) DOES NOT ELONGATE THE AREA WHERE THE DISK AND THE DYNAMIZATION PARTS ARE. THE EXTERNAL FIXATOR (MEFISTO SYSTEM) WAS PLACED SIX DAYS PRIOR TO (B)(6) 2016. THE REPORTED ISSUE WAS OBSERVED ON (B)(6) 2016; ON THE SIXTH DAY OF THE ELONGATION TREATMENT. THE SURGEON TURNED ON THE DISC HOLE TO TEST THE ELONGATION FUNCTION BUT IT APPEARED THAT THE SYSTEM (O-RING) IS BLOCKED AND CANNOT ROTATE. THE MEFISTO SYSTEM WAS REMOVED ON (B)(6) 2016 DUE TO THE CENTRAL BODY'S INABILITY TO PERFORM ELONGATION. THIS REPORT IS 1 OF 3 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
142872 CENTRAL BODY F/MEF MAX-L 437 MIN-L 287 TRACTION, APPARATUS, NON-POWERED HST SYNTHES HAGENDORF 8284785

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention