HLM TUBING SET W/SOFTLINE COATING
Report
- Report Number
- 8010762-2016-00134
- Event Type
- Malfunction
- Date Received
- March 3, 2016
- Date of Event
- February 9, 2016
- Report Date
- February 9, 2016
- Manufacturer
- MAQUET CARDIOPULMONARY AG
- Product Code
- DWE
- PMA / PMN Number
- K082117
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). THE PRODUCT WAS DISPOSED OF BY THE HOSPITAL; THEREFORE, IT IS NOT AVAILABLE FOR MANUFACTURER'S LABORATORY INVESTIGATION. MAQUET CARDIOPULMONARY IS AWARE OF A SIMILAR COMPLAINT WHICH WAS INVESTIGATED WITH THE FOLLOWING RESULTS: THE PRODUCT WAS TESTED IN THE LABORATORY OF MAQUET CARDIOPULMONARY FOR ITS DE-AIRING BEHAVIOR (NO AIR ENTRY THROUGH DE-AIRING MEMBRANE) DURING PRIMING ACCORDING TOLV 974 AND IFU QUADROX-I ADULT G-117 VERSION 06. DURING THE WHOLE TESTING PROCESS NO AIR BUBBLE CREATION AND NO AIR ACCUMULATION WAS DETECTED. CONCLUSION: THE FAILURE "AIR COMES THROUGH HYDROPHOBIC FILTER" WAS NOT CONFIRMED. BASED ON THE ABOVE MENTIONED INVESTIGATION RESULTS OF THE SIMILAR COMPLAINT A CONFIRMATION OF THE FAILURE IS NOT POSSIBLE. A REVIEW OF THE DEVICE HISTORY RECORDS OF THE OXYGENATOR WAS NOT POSSIBLE AS THE PRODUCT WAS DISPOSED OF AND THE UDI IS UNKNOWN. ADDITIONAL INFORMATION: THE PRODUCT MENTIONED UNDER SECTION D IS A TUBING SET AND THE INCLUDED AFFECTED COMPONENT HAS THE CONTRIBUTING DESIGN FUNCTION OF THE QUADROX-I WHICH IS REGISTERED UNDER 510(K): K082117.
IT WAS REPORTED THE HYDROPHOBIC FILTER IS DEFECTIVE: THIS FILTER ALLOWS AN ENTRY OF AIR GOING TO THE OXYGENATOR DURING THE PRIMING PROCEDURE WITH 2.4L/MIN. THE AIR ENTRY IS OBSERVED WHEN THE RATE IS CHANGED FROM 2.4 TO 0.7L/MIN DURING A PERIOD GOING FROM 1 TO 2 SECONDS ( BUT NOT BY USING 7L/MIN WITH ON/OFF). (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 133518 | HLM TUBING SET W/SOFTLINE COATING | TUBING, PUMP, CARDIOPULMONARY BYPASS | DWE | MAQUET CARDIOPULMONARY AG | HQV 69001 | 92166861 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |