FDA Adverse Event Malfunction Summary report: N

HLM TUBING SET W/SOFTLINE COATING

MDR report key: 5477315 · Received March 3, 2016

Report

Report Number
8010762-2016-00134
Event Type
Malfunction
Date Received
March 3, 2016
Date of Event
February 9, 2016
Report Date
February 9, 2016
Manufacturer
MAQUET CARDIOPULMONARY AG
Product Code
DWE
PMA / PMN Number
K082117
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE PRODUCT WAS DISPOSED OF BY THE HOSPITAL; THEREFORE, IT IS NOT AVAILABLE FOR MANUFACTURER'S LABORATORY INVESTIGATION. MAQUET CARDIOPULMONARY IS AWARE OF A SIMILAR COMPLAINT WHICH WAS INVESTIGATED WITH THE FOLLOWING RESULTS: THE PRODUCT WAS TESTED IN THE LABORATORY OF MAQUET CARDIOPULMONARY FOR ITS DE-AIRING BEHAVIOR (NO AIR ENTRY THROUGH DE-AIRING MEMBRANE) DURING PRIMING ACCORDING TOLV 974 AND IFU QUADROX-I ADULT G-117 VERSION 06. DURING THE WHOLE TESTING PROCESS NO AIR BUBBLE CREATION AND NO AIR ACCUMULATION WAS DETECTED. CONCLUSION: THE FAILURE "AIR COMES THROUGH HYDROPHOBIC FILTER" WAS NOT CONFIRMED. BASED ON THE ABOVE MENTIONED INVESTIGATION RESULTS OF THE SIMILAR COMPLAINT A CONFIRMATION OF THE FAILURE IS NOT POSSIBLE. A REVIEW OF THE DEVICE HISTORY RECORDS OF THE OXYGENATOR WAS NOT POSSIBLE AS THE PRODUCT WAS DISPOSED OF AND THE UDI IS UNKNOWN. ADDITIONAL INFORMATION: THE PRODUCT MENTIONED UNDER SECTION D IS A TUBING SET AND THE INCLUDED AFFECTED COMPONENT HAS THE CONTRIBUTING DESIGN FUNCTION OF THE QUADROX-I WHICH IS REGISTERED UNDER 510(K): K082117.

Description of Event or Problem · 1

IT WAS REPORTED THE HYDROPHOBIC FILTER IS DEFECTIVE: THIS FILTER ALLOWS AN ENTRY OF AIR GOING TO THE OXYGENATOR DURING THE PRIMING PROCEDURE WITH 2.4L/MIN. THE AIR ENTRY IS OBSERVED WHEN THE RATE IS CHANGED FROM 2.4 TO 0.7L/MIN DURING A PERIOD GOING FROM 1 TO 2 SECONDS ( BUT NOT BY USING 7L/MIN WITH ON/OFF). (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
133518 HLM TUBING SET W/SOFTLINE COATING TUBING, PUMP, CARDIOPULMONARY BYPASS DWE MAQUET CARDIOPULMONARY AG HQV 69001 92166861

Patients

Seq Age Sex Outcome Treatment
1