FDA Adverse Event Other Summary report: N

CHEMO-PORT

MDR report key: 547655 · Received October 7, 2004

Report

Report Number
2925153-2004-00020
Event Type
Other
Date Received
October 7, 2004
Date of Event
September 24, 2004
Report Date
September 27, 2004
Manufacturer
HDC CORP.
Product Code
LJT
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

"A PT HAD A PORT INSERTED 6 WEEKS AGO. TODAY PT PRESENTED WITH THE DISTAL 10CM FRACTURED OFF AND LODGED IN THE RIGHT VENTRICLE WITH THE REMAINING PORT AND LINE IN PLACE. PT DID HAVE A RATHER NARROW SPACE BETWEEN THE FIRST AND CLAVICLE ON INSERTION, AND DR PRESUMES THE SCISSORS ACTION OF THE TWO AGAINST THE LINE PORTION OF THE PORT, CUT IT IN TWO AND IT FLOATED DOWNSTREAM. THE SEVERED END OF THE PORT WAS REMOVED BY CATHETER SNARE IN AN ANGIOGRAPHY SUITE. THE PORT WILL NOT BE REMOVED FULLY UNTIL THE BLEEDING RISK IS REDUCED."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CHEMO-PORT * LJT HDC CORP. NA NA

Patients

Seq Age Sex Outcome Treatment
1 * Other