FDA Adverse Event Other Summary report: N

NARKOMED GS ANESTHESIA MACHINE

MDR report key: 547596 · Received September 27, 2004

Report

Report Number
547596
Event Type
Other
Date Received
September 27, 2004
Date of Event
September 27, 2004
Report Date
September 27, 2004
Manufacturer
DRAEGER MEDICAL, INC
Product Code
BSZ
Product Problem
Yes
Report Source
User Facility report
Reporter Location
AZ, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

AFTER INTUBATING THE PATIENT, THE BELLOWS ON THE ANESTHESIA MACHINE VENTILATOR STUCK IN THE DOWN POSITION. THE DOCTOR PROCEEDED TO 'BAG' THE PATIENT AND THEN WAS ABLE TO GET THE BELLOWS OPERATING AGAIN BY ADJUSTING THE INSPIRATIONAL PRESSURE LIMIT ON THE MACHINE. NO NEGATIVE PATIENT OUTCOME. COMPANY NOTIFIED AND WILL COME OUT TO CHECK THE ANESTHESIA MACHINE; MACHINE IS UNDER WARRANTY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NARKOMED GS ANESTHESIA MACHINE ANESTHESIA MACHINE BSZ DRAEGER MEDICAL, INC * *

Patients

Seq Age Sex Outcome Treatment
1 83 YR