DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2016-01341
- Event Type
- Malfunction
- Date Received
- March 1, 2016
- Date of Event
- February 3, 2016
- Report Date
- February 3, 2016
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- UDI-DI
- 00386270000224
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4) DESCRIBE EVENT OR PROBLEM - ADDITIONAL, DEVICE AVAILABLE FOR EVALUATION - ADDITIONAL, ADDITIONAL INFORMATION/DEVICE EVALUATION, DEVICE EVALUATED BY MANUFACTURER - ADDITIONAL, EVENT PROBLEM AND EVALUATION CODES - ADDITIONAL.
(B)(4)
THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE TEST FAILED. THE CUSTOMER COMPLAINT WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE TRANSMITTER.
PATIENT CONTACTED DEXCOM ON (B)(6) 2016 TO REPORT THAT ON (B)(6) 2016, THE G5 TRANSMITTER WOULD NOT PAIR WITH THE G5 APPLICATION. PATIENT CONFIRMED SMART DEVICE IS NOT IN AIRPLANE MODE AND WI-FI IS ENABLED. PATIENT STATED THEY RE-INSTALLED G5 APPLICATION AND RESTARTED SMART DEVICE, BUT IT IS STILL NOT CONNECTING. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 127734 | DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | 9438-06 | 5206846 | 00386270000224 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR |