FDA Adverse Event Malfunction Summary report: N

STELLANT CT SYRINGE KIT

MDR report key: 5470998 · Received March 1, 2016

Report

Report Number
2520313-2016-00010
Event Type
Malfunction
Date Received
March 1, 2016
Date of Event
February 3, 2016
Report Date
February 3, 2016
Manufacturer
BAYER MEDICAL CARE, INC.
Product Code
DXT
PMA / PMN Number
K082905
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

BAYER QUALITY ASSURANCE PRODUCT ANALYSIS RECEIVED AND EXAMINED A PHOTOGRAPH OF THE TUBING, LOT # 162065. THE ACTUAL PRODUCT WAS NOT AVAILABLE FOR EXAMINATION. VISUAL INSPECTION OF THE PHOTOGRAPH SHOWS THE PARTICULATE IS PARTIALLY EMBEDDED WITH EXTENSIONS WITHIN THE FLUID PATH. THIS PARTICLE WAS INTRODUCED DURING COMPONENT MANUFACTURING. THE SYRINGE KIT INSTRUCTIONS FOR USE INSTRUCTS THE USER TO "VISUALLY INSPECT CONTENTS AND PACKAGE BEFORE EACH USE". THIS VISUAL INSPECTION IS TO ENSURE PARTICULATES ARE NOTICED AND MITIGATED.

Additional Manufacturer Narrative · 1

BAYER QUALITY ASSURANCE PRODUCT ANALYSIS RECEIVED AND EXAMINED THE RETURNED LOW PRESSURE CONNECTING TUBING SET (LPCT), LOT # 162065. THE DEFECT WAS IDENTIFIED AS DEGRADED PLASTIC COMPOUND THAT IS PARTIALLY EMBEDDED IN THE TUBING WALL AND EXTENDS INTO THE FLUID PATHWAY. DYNAMIC FLOW STUDIES WERE PERFORMED ON THE TUBING WITH BOTH SALINE AND CONTRAST MEDIA, CONCLUDING THAT THE EMBEDDED COMPOUND COULD NOT DISLODGE OR FRAGMENT INTO THE FLUID PATH. THE CAUSE OF THE REPORTED PROBLEM IS DEGRADED COMPOUND (RESIN) THAT COLLECTED ON THE EXTRUSION TOOLING AND BECAME PARTIALLY EMBEDDED IN THE TUBING AS IT WAS FORMED DURING MANUFACTURING.

Description of Event or Problem · 1

THE SITE REPORTED THE FOLLOWING: A BLACK FOREIGN PARTICLE WAS DETECTED INSIDE THE TUBING OF A STELLANT SYRINGE KIT. THIS WAS DETECTED BY THE STAFF AFTER FILLING THE SYRINGE WITH CONTRAST MEDIA AND INJECTING INTO THE PATIENT.