FDA Adverse Event Malfunction Summary report: N

AISYS CS2

MDR report key: 5470355 · Received March 1, 2016

Report

Report Number
2112667-2016-00398
Event Type
Malfunction
Date Received
March 1, 2016
Date of Event
February 9, 2016
Report Date
March 1, 2016
Manufacturer
DATEX-OHMEDA, INC.
Product Code
BSZ
PMA / PMN Number
K132530
Removal / Correction Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE DRIVE GAS CHECK VALVE COULD BECOME STUCK IN A FIXED OPEN POSITION WHICH COULD CAUSE PRESSURE TO BUILD IN THE MECHANICAL VENTILATION CYCLE. IF THIS ISSUE IS LEFT UNRESOLVED, IT COULD RESULT IN EXCESSIVE OR PROLONGED PRESSURE IN THE PATIENT BREATHING CIRCUIT DURING VENTILATION POTENTIALLY RESULTING IN BAROTRAUMA. GE HEALTHCARE (GEHC) REPORTED A FIELD MODIFICATION FOR THIS ISSUE PER 21 CFR 806 ON (B)(6) 2015. THE RECALL CLASSIFICATION NUMBER IS Z-0677-2016 THROUGH Z-0682-2016. NO REPORT OF PATIENT INVOLVEMENT. THE INITIAL REPORTER IS LOCATED OUTSIDE THE U.S., AND THEREFORE THIS INFORMATION IS NOT PROVIDED DUE TO COUNTRY PRIVACY LAWS. THE DRIVE GAS CHECK VALVE COULD BECOME STUCK IN A FIXED OPEN POSITION WHICH COULD CAUSE PRESSURE TO BUILD IN THE MECHANICAL VENTILATION CYCLE. IF THIS ISSUE IS LEFT UNRESOLVED, IT COULD RESULT IN EXCESSIVE OR PROLONGED PRESSURE IN THE PATIENT BREATHING CIRCUIT DURING VENTILATION POTENTIALLY RESULTING IN BAROTRAUMA. GE HEALTHCARE (GEHC) REPORTED A FIELD MODIFICATION FOR THIS ISSUE PER 21 CFR 806 ON (B)(6) 2015. THE RECALL CLASSIFICATION NUMBER IS Z-0677-2016 THROUGH Z-0682-2016. NO REPORT OF PATIENT INVOLVEMENT. THE INITIAL REPORTER IS LOCATED OUTSIDE THE U.S., AND THEREFORE THIS INFORMATION IS NOT PROVIDED DUE TO COUNTRY PRIVACY LAWS.

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT, DURING A PREOPERATIVE CHECKOUT, POSITIVE END EXPIRATORY PRESSURE WAS HIGHER THAN EXPECTED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
128644 AISYS CS2 ANESTHESIA GAS MACHINE BSZ DATEX-OHMEDA, INC.

Patients

Seq Age Sex Outcome Treatment
1