IMMULITE 1000 IGF-1
Report
- Report Number
- 2432235-2016-00109
- Event Type
- Malfunction
- Date Received
- February 29, 2016
- Date of Event
- November 24, 2015
- Report Date
- February 4, 2016
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- CFL
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE CUSTOMER CONTACTED A SIEMENS CUSTOMER CARE CENTER (CCC) SPECIALIST. A SIEMENS DESIGNATED COMPLAINT HANDLING UNIT (DCU) DETERMINED THAT IGF-1 REAGENT LOTS ARE KITTED AND VALIDATED WITH SPECIFIC PRE-TREATMENT DILUENT LOTS. IGF-1 REAGENT LOT 412 WAS KITTED AND VALIDATED WITH PRE-TREATMENT DILUENT LOT 057. NO OTHER LOT OF PRE-TREATMENT DILUENTS CAN BE USED WITH IGF-1 REAGENT LOT 412. THE CAUSE OF INCORRECT USAGE OF PRE-TREATMENT DILUENT LOT 055 WITH IGF-1 REAGENT LOT 412 IS DUE TO OFF-LABEL USE.
THE OPERATOR OF AN IMMULITE 1000 INSTRUMENT REPORTED THAT WHILE USING INSULIN-LIKE GROWTH FACTOR 1 (IGF-1) REAGENT LOT 412 IN COMBINATION WITH PRE-TREATMENT DILUENT LOT 055, IT TAKES A LONG TIME BEFORE THE IGF-1 IS RELEASED FROM BINDING PROTEIN 3 (BP3). THE CUSTOMER STATED THAT IT TAKES APPROXIMATELY 2 HOURS OF PRE-INCUBATION BEFORE A STABLE IGF-1 RESULT IS OBTAINED. THE CUSTOMER ALSO STATED THAT WHILE RUNNING TESTS WITH PRE-TREATMENT DILUENT LOT 057 DID NOT SHOW THIS PHENOMENON. THE CUSTOMER WAS NOT REPORTING ANY PATIENT RESULTS WITH REAGENT LOT 412 AND PRE-TREATMENT DILUENT LOT 055. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE USE OF INCORRECT COMBINATION OF PRE-TREATMENT DILUENT LOT 055 WITH IGF-1 REAGENT LOT 412.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 123267 | IMMULITE 1000 IGF-1 | IMMULITE 1000 IGF-1 | CFL | SIEMENS HEALTHCARE DIAGNOSTICS INC. | IMMULITE 1000 IGF-1 | 412 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |