DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM
Report
- Report Number
- 3004753838-2016-41321
- Event Type
- Malfunction
- Date Received
- February 29, 2016
- Date of Event
- February 3, 2016
- Report Date
- February 3, 2016
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- UDI-DI
- 00386270000156
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
(B)(4).
(B)(4).
(B)(4).
PATIENT CONTACTED DEXCOM ON (B)(6) 2016 TO REPORT THAT ON (B)(6) 2016, THE PATIENT EXPERIENCED A PERMANENT OUT OF RANGE SIGNAL. THE COMPLAINT DEVICE WAS NOT RETURNED FOR EVALUATION. HOWEVER, THE DATA LOG WAS PROVIDED BY THE PATIENT. THE DATA WAS REVIEWED ON (B)(6) 2016 AND COULD NOT CONFIRM THE REPORTED LOSS OF CONNECTION. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE.
IN ADDITION TO THE COMPLAINT RECEIVER, THE TRANSMITTER (PART NUMBER STT-GL-003/SERIAL NUMBER (B)(4)/LOT NUMBER 5202766) WAS RETURNED ON 06/01/2016. THE RETURNED TRANSMITTER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND FAILED TESTING. THE REPORTED EVENT OF UNRECOVERABLE LOSS OF ANTENNA PASSED THRESHOLD WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE TRANSMITTER.
THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THERE WAS NO FAILURE DETECTED RELATED TO THE COMPLAINT. THE RECEIVER LOG WAS REVIEWED AND NO ERRORS WERE OBSERVED. THE REPORTED EVENT OF A PERMANENT OUT OF RANGE SIGNAL WAS NOT CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 125402 | DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM | MDS | MDS | DEXCOM, INC. | MT22495 | 5203315 | 00386270000156 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR |