FDA Adverse Event Malfunction Summary report: N

ALERE DETERMINE HIV-1/2 AG/AB COMBO

MDR report key: 5465993 · Received February 29, 2016

Report

Report Number
3008754095-2016-00005
Event Type
Malfunction
Date Received
February 29, 2016
Date of Event
December 1, 2015
Report Date
February 5, 2016
Manufacturer
ORGENICS LTD
Product Code
MZF
UDI-DI
07290015003735
PMA / PMN Number
120037/0
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT INVOLVES REPORTED (B)(6) RESULT OF THE AB LINE. FOLLOWING THE (B)(6) RESULT, THE PATIENT TESTED (B)(6) . AN INVESTIGATION HAS BEEN CONDUCTED BY ORGENICS WITH THE FOLLOWING ACTIVITIES AND RESULTS: - THE ASSAY PERFORMANCE WAS ASSESSED BY TESTING A KIT LOT # 150512 FROM QC RETENTION. THE KIT PERFORMED ACCORDINGLY WITH QC SPECIFICATION. - HISTORY RECORD OF RELEASE TESTING (QC) DATA AND BATCH RECORDS FOR LOT #150512 WERE REVIEWED. ALL QUALITY CONTROL RELEASE TESTING WAS VALID AND PERFORMED WITHIN SPECIFICATIONS WITH NO OBSERVED ANOMALIES NOTED BASED ON THE RESULTS OF THE INVESTIGATION A DEFINITIVE ROOT CAUSE OF THE COMPLAINT COULD NOT BE DETERMINED. ONE POTENTIAL CAUSE IN THIS CASE COULD BE THE PRESENCE OF A SUBSTANCE WITHIN THE SAMPLE THAT COULD INFLUENCE THE TEST RESULT, SUCH AS TOXOPLASMA IGG, HUMAN ANTI-MOUSE ANTIBODIES, RHEUMATOID FACTOR, ELEVATED TRIGLYCERIDES, OR HERPES SIMPLEX VIRUS INFECTION. AS STATED IN THE WARNINGS SECTION OF THE PRODUCT INSERT FOR THE ALERE DETERMINE (B)(6) COMBO: "SPECIMENS FROM INDIVIDUALS WITH TOXOPLASMA IGG, HUMAN ANTI-MOUSE ANTIBODIES, RHEUMATOID FACTOR, ELEVATED TRIGLYCERIDES (ABOVE 600 MG/DL), HERPES SIMPLEX VIRUS INFECTION, HOSPITALIZED AND CANCER PATIENTS MAY GIVE (B)(6) TEST RESULTS." ORGENICS HAS INITIATED A CORRECTIVE ACTION / PREVENTIVE ACTION (CAPA) INVESTIGATION (INTERNAL REFERENCE CAPA(B)(4)) TO FURTHER INVESTIGATE THE CAUSE OF THIS EVENT.

Additional Manufacturer Narrative · 1

CORRECTIVE ACTION / PREVENTIVE ACTION (CAPA) INVESTIGATION (INTERNAL REFERENCE (B)(4)) HAS BEEN COMPLETED FOR FURTHER INVESTIGATION OF REOCCURRING FALSE POSITIVE RESULTS WHEN TESTING LABOR AND DELIVERY SAMPLES. (B)(4) ASSESSED THE SIGNIFICANCE OF THE FALSE POSITIVE RATES TO THE SAFETY AND EFFECTIVENESS OF THE TEST IN PREGNANT AND LABOR AND DELIVERY PATIENTS. INVESTIGATION CONCLUSIONS MADE WITHIN THE CAPA INDICATE THAT ALERE DETERMINE (B)(6) COMBO PRODUCES RESULTS WITHIN THE RANGE EXPECTED AS MENTIONED IN THE PRODUCT PI AND HAS PERFORMED SIMILARLY WITH ANOTHER FDA-APPROVED 4TH GENERATION TEST IN A COMPARABLE FIELD STUDY. BASED ON THE ABOVE, THERE IS NO EVIDENCE TO SUGGEST A PRODUCT DEFICIENCY AND THE INVESTIGATION IS DEEMED CLOSED. THE TREND WILL BE FURTHER MONITORED AND TRACKED.

Description of Event or Problem · 1

CUSTOMER REPORTED A POTENTIAL (B)(6) RESULT FOR A LABOR AND DELIVERY PATIENT. CONFIRMATORY TESTING WAS PERFORMED AND THE RESULT WAS (B)(6). THERE WERE NO ADVERSE PATIENT OUTCOMES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
123606 ALERE DETERMINE HIV-1/2 AG/AB COMBO IVD FOR HIV AG/AB MZF ORGENICS LTD 150512 07290015003735

Patients

Seq Age Sex Outcome Treatment
1 39 YR