FDA Adverse Event Other Summary report: N

EMERSON

MDR report key: 545549 · Received September 17, 2004

Report

Report Number
1216146-2004-00002
Event Type
Other
Date Received
September 17, 2004
Date of Event
October 17, 2003
Report Date
August 18, 2004
Manufacturer
J. H. EMERSON CO.
Product Code
FMT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

IT WAS DECIDED AT A FOLLOW-UP MEETING TO THIS EVENT TO SUBMIT A VOLUNTARY MEDWATCH REPORT. MODERATE BURNS WERE CAUSED BY A HEAT THERAPY LAMP DETERMINED TO BE TOO CLOSE TO THE PT. WHEN THE LAMP WAS EXAMINED AT THE TIME OF THE INCIDENT, IT WAS DISCOVERED THAT THE SLEEVE CAP SURROUNDING THE HEIGHT ADJUSTMENT SLIDE HAD BROKEN OFF AND WAS MISSING. THIS ALLOWS FREE PLAY WHICH COULD HAVE ALLOWED THE LAMP TO SLIDE DOWN TO THE LEVEL WHERE THE SAFETY PIN WAS ENGAGED. (2-4 IN. ). IT WAS UNDETERMINED IF THIS WAS A CONTRIBUTING FACTOR IN THIS INCIDENT, HOWEVER USERS SHOULD BE WARNED TO MONITOR CORRECT DISTANCE (28") BETWEEN LAMP AND PT. ALSO, SLEEVE CAPS WHICH PREVENT HEIGHT ADJUSTMENT FREEPLAY SHOULD BE EXAMINED FOR CRACKS DURING ROUTINE SERVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EMERSON WARMING LIGHT FMT J. H. EMERSON CO. 96-TDL *

Patients

Seq Age Sex Outcome Treatment
1 4 MO Other