FDA Adverse Event Malfunction Summary report: N

BARIATRIC BED HEAD 9153641516

MDR report key: 5451563 · Received February 22, 2016

Report

Report Number
1031452-2016-00626
Event Type
Malfunction
Date Received
February 22, 2016
Date of Event
February 8, 2016
Report Date
February 8, 2016
Manufacturer
INVACARE FLORIDA OPERATIONS
Product Code
OSI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A SUPPLEMENTAL RECORD WILL BE FILED.

Additional Manufacturer Narrative · 1

THIS EVENT IS A DUPLICATE OF REPORT NUMBER 1031452-2016-00625. IT CONCERNS BAR5000IVC (HEAD SECTION) WITH SERIAL NUMBER (B)(4), WHICH IS FROM THE SAME BED (MODEL BAR600IVC) AS BAR5490IVC (FOOT SECTION) WITH SERIAL NUMBER (B)(4) REFERENCED IN 1031452-2016-00625.

Description of Event or Problem · 1

THE DEALER STATED THAT THEY RECEIVED A CORRESPONDENCE FROM THE VA STATING: EMERGENCY: PLEASE ASSESS/REPAIR CONTROLS ON HOSPITAL BED. PATIENT REPORTS THE WIRES BURNED UP, CAUSING THE FIRE DEPT. TO BE CALLED, AND THE BED WILL NOT WORK, IS STUCK WITH THE HEAD UP , PLACING THE PATIENT IN AN AWKWARD/DANGEROUS POSITION.

Description of Event or Problem · 1

THE DEALER STATED THAT THEY RECEIVED A CORRESPONDENCE FROM THE VA STATING: EMERGENCY: PLEASE ASSESS/REPAIR CONTROLS ON HOSPITAL BED. PATIENT REPORTS THE WIRES BURNED UP, CAUSING THE FIRE DEPT. TO BE CALLED, AND THE BED WILL NOT WORK, IS STUCK WITH THE HEAD UP , PLACING THE PATIENT IN AN AWKWARD/DANGEROUS POSITION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
109749 BARIATRIC BED HEAD 9153641516 BARIATRIC BED OSI INVACARE FLORIDA OPERATIONS BAR600IVC

Patients

Seq Age Sex Outcome Treatment
1