FDA Adverse Event Injury Summary report: N

GORE® TAG® THORACIC ENDOPROSTHESIS

MDR report key: 5451379 · Received February 22, 2016

Report

Report Number
2017233-2016-00146
Event Type
Injury
Date Received
February 22, 2016
Date of Event
January 17, 2012
Report Date
January 25, 2016
Manufacturer
W.L. GORE & ASSOCIATES
Product Code
MIH
PMA / PMN Number
P040043
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL MANUFACTURER NARRATIVE - THE REVIEW OF THE MANUFACTURING PAPERWORK VERIFIED THAT THIS LOT MET ALL PRE-RELEASE SPECIFICATIONS. (B)(4).

Description of Event or Problem · 1

ON (B)(6) 2010, THIS PATIENT UNDERWENT ENDOVASCULAR REPAIR OF A DESCENDING THORACIC AORTIC ANEURYSM USING TWO GORE® TAG® THORACIC ENDOPROSTHESES (TGT3415/7064618-DISTAL, TGT3420/7160084-PROXIMAL). THE PREOPERATIVE ANEURYSM DIAMETER MEASURED 81 MM. THE DIAMETER OF THE DISTAL NECK MEASURED 24-29 MM. ON (B)(6) 2010, THE PATIENT WAS DISCHARGED FROM THE HOSPITAL. ON (B)(6) 2010, FOLLOW-UP IMAGES SHOWED NO ENDOLEAK. THE ANEURYSM DIAMETER MEASURED 76 MM. ON (B)(6) 2010, FOLLOW-UP IMAGES SHOWED THE ANEURYSM DIAMETER MEASURED 76 MM. ON (B)(6) 2011, FOLLOW-UP IMAGES SHOWED THE ANEURYSM DIAMETER MEASURED 77 MM. ON (B)(6) 2012, FOLLOW-UP IMAGES SHOWED A DISTAL TYPE I ENDOLEAK. THE ANEURYSM DIAMETER MEASURED 82 MM. A GORE® TAG® THORACIC ENDOPROSTHESIS (TGT3420/9119484) WAS IMPLANTED TO TREAT THE ENDOLEAK. THE CAUSE OF THE ENDOLEAK IS UNKNOWN. ON (B)(6) 2012, FOLLOW-UP IMAGES SHOWED A TYPE II ENDOLEAK. THE ANEURYSM DIAMETER MEASURED 89 MM. ON (B)(6) 2013, THE ANEURYSM DIAMETER MEASURED 98 MM. ON (B)(6) 2015, THE ANEURYSM DIAMETER MEASURED 109 MM. NO FURTHER REINTERVENTIONS HAVE BEEN REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
107666 GORE® TAG® THORACIC ENDOPROSTHESIS SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH W.L. GORE & ASSOCIATES 7064618

Patients

Seq Age Sex Outcome Treatment
1 80 YR Hospitalization| R TGT3420/71600841/17/12: TGT3420/9119484