FAST-FIX 360 CURVED NDL DELIVERY SYS
Report
- Report Number
- 1219602-2016-00095
- Event Type
- Malfunction
- Date Received
- February 22, 2016
- Date of Event
- February 10, 2016
- Report Date
- February 18, 2016
- Manufacturer
- SMITH & NEPHEW, INC.
- Product Code
- GAT
- PMA / PMN Number
- K092508
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
EVALUATION NARRATIVE - ONE FAST-FIX 360 CURVED NEEDLE DELIVERY SYSTEM LOT 5051756 WERE RETURNED FOR EVALUATION. VISUAL ASSESSMENT OF THE DEVICE SHOWS THE INSERTION NEEDLE IS TWISTED AND IS BENT ON TWO PLANES. DEVICE WAS FUNCTIONALLY TESTED FOR PROPER ACTUATOR ADVANCEMENT AND CYCLING, DEVICE WOULD NOT FUNCTION PROPERLY DUE TO THE AFOREMENTIONED DAMAGE. NO TS OR SUTURE WERE RETURNED FOR EXAMINATION. REPORTED NON-DEPLOYMENT COULD NOT BE CONFIRMED. PER THE DEVICE IFU UNDER WARNINGS ¿DO NOT BEND THE DELIVERY NEEDLE. THE FAST-FIX 360 DEVICES ARE MANUFACTURED WITH STRAIGHT OR CURVED DELIVERY NEEDLES. INTENTIONAL BENDING OF THE DELIVERY NEEDLE MAY MAKE IT DIFFICULT OR IMPOSSIBLE TO DELIVER THE T1 AND T2 IMPLANTS. IF THE DELIVERY NEEDLE HAS BEEN INADVERTENTLY BENT, OR IF RESISTANCE IS ENCOUNTERED DURING DEPLOYMENT, A NEW DELIVERY DEVICE MAY BE NEEDED¿. NO ROOT CAUSE RELATED TO THE MANUFACTURE OF THE DEVICE CAN BE ESTABLISHED. NO FURTHER INVESTIGATION IS WARRANTED AT THIS TIME. (B)(4).
(B)(4).
IT WAS REPORTED THAT THE FAST-FIX 360 T1 IMPLANT FIRED BUT T2 DID NOT. A BACKUP DEVICE WAS REPORTEDLY USED TO COMPLETE THE PROCEDURE WITH A FIVE MINUTE DELAY. THE CUSTOMER REPORTED THAT THEY WERE ABLE TO COMPLETELY REMOVE THE ANCHORS AND THEY DID NOT HAVE TO MAKE AN ADDITIONAL INCISION." THERE IS NO REPORT OF PATIENT INJURY AS A RESULT OF THE ALLEGED EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 107734 | FAST-FIX 360 CURVED NDL DELIVERY SYS | SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE | GAT | SMITH & NEPHEW, INC. | 50571756 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |