FDA Adverse Event Malfunction Summary report: N

ENDO CLIP APPLIER

MDR report key: 5449179 · Received February 19, 2016

Report

Report Number
3005075853-2016-01076
Event Type
Malfunction
Date Received
February 19, 2016
Date of Event
January 27, 2016
Report Date
February 5, 2016
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). BATCH # UNK. THE LOT HISTORY RECORDS WERE REVIEWED AND THE MANUFACTURING CRITERIA WERE MET PRIOR TO THE RELEASE OF THIS LOT. ATTEMPTS HAVE BEEN MADE TO RETRIEVE THE DEVICE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. THE LOT HISTORY RECORDS WERE REVIEWED AND THE MANUFACTURING CRITERIA WERE MET PRIOR TO THE RELEASE OF THIS LOT. HOW DID CLIPS NOT LOAD PROPERLY? DID CLIPS ONLY SLOW FEED? OR DID CLIPS ALSO SIDEWAYS FEED? OR DID MULTIPLE CLIPS FEED? OR DID DEVICE PARTIALLY FEED CLIPS? OR DID CLIPS DROP AND EJECT?

Additional Manufacturer Narrative · 1

(B)(4). BATCH # = K92029. THE ANALYSIS RESULTS FOUND THAT THE ER420 INSTRUMENT WAS RECEIVED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS FUNCTIONALLY EVALUATED. DURING THE ANALYSIS, THE DEVICE WAS CYCLED AND THE REMAINING 12 CLIPS WERE EJECTED; FINALLY THE INSTRUMENT LOCKED OUT AS INTENDED. NO SLOW FEEDING ISSUES WERE NOTED DURING ANALYSIS. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE DEVICE WAS LOADING EXTREMELY SLOW. THE CLIPS WERE ALSO NOT LOADING PROPERLY FORMING MALFORMED CLIPS. IT IS UNKNOWN HOW THE CASE WAS COMPLETED. THERE WERE NO PATIENT CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
106005 ENDO CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA K4D89R

Patients

Seq Age Sex Outcome Treatment
1