Report Number
0002031963-2016-00002
Event Type
Injury
Date Received
February 19, 2016
Date of Event
January 22, 2016
Report Date
April 20, 2020
Manufacturer
BERCHTOLD GMBH & CO. KG
Product Code
FQO
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT CODE FOR THE INITIAL FILING WAS FILED INCORRECTLY. THIS RECORD HAS BEEN UPDATED TO REFLECT THE CORRECT PRODUCT CODE OF FQO.

Additional Manufacturer Narrative · 1

IT WAS REPORTED THAT IN GI ROOM 7 THE MEDIUM DUTY SUSPENSION WITH DUAL FLAT PANEL MONITOR (DFPM) ALLEGEDLY DRIFTED. REPORTEDLY, THE NURSE HIT THE DFPM UPON STANDING UP AFTER BENDING OVER TO PICK SOMETHING UP OFF OF THE FLOOR. ALLEGEDLY, THE NURSE HAD SIGNIFICANT HEAD PAIN AND WAS SEEN IN THE EMERGENCY DEPARTMENT. IT WAS REPORTED THAT THE EMERGENCY DEPARTMENT PERFORMED A NEURO ASSESSMENT. THE NURSE PASSED THE ASSESSMENT AND WAS DISCHARGED. THE NURSE REPORTEDLY RETURNED TO THE EMERGENCY DEPARTMENT TWO DAYS LATER TO BE REASSESSED AS SHE REPORTED COMPLAINTS OF A SEVERE HEADACHE AND VOMITING. THE EMERGENCY DEPARTMENT PERFORMED A CT SCAN ON HER HEAD, WHICH WAS NEGATIVE. AFTER FURTHER INVESTIGATION, A STRYKER FIELD SERVICE TECHNICIAN (SFST) PERFORMED A ROTATION TEST AS WELL AS A SPRING ARM ARTICULATION TEST AND WAS UNABLE TO DUPLICATE THE COMPLAINT. THERE WAS NO CONFIRMED MALFUNCTION OF THE DEVICE AND NO CONFIRMATION OF PERMANENT IMPAIRMENT OR DAMAGE TO THE USER.

Description of Event or Problem · 1

IT WAS REPORTED THAT IN GI ROOM 7 THE DUAL FLAT PANEL MONITOR (DFP) ALLEGEDLY DRIFTED AND THAT A NURSE HIT THE DFP UPON STANDING AFTER BENDING OVER TO PICK SOMETHING OFF OF THE FLOOR. SHE REPORTED TO HAVE SIGNIFICANT HEAD PAIN AND WENT TO THEIR EMERGENCY DEPARTMENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT IN GI ROOM 7 THE DUAL FLAT PANEL MONITOR (DFP) ALLEGEDLY DRIFTED AND THAT A NURSE HIT THE DFP UPON STANDING AFTER BENDING OVER TO PICK SOMETHING OFF OF THE FLOOR. SHE REPORTED TO HAVE SIGNIFICANT HEAD PAIN AND WENT TO THEIR EMERGENCY DEPARTMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
104513 MEDIUM DUTY DUAL YOKE MEDIUM DUTY DUAL YOKE FQO BERCHTOLD GMBH & CO. KG

Patients

Seq Age Sex Outcome Treatment
1