TRUERESULT
Report
- Report Number
- 1052693-2016-00306
- Event Type
- Malfunction
- Date Received
- February 18, 2016
- Date of Event
- January 26, 2016
- Report Date
- February 18, 2016
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL
- Reporter Occupation
- OTHER
Narratives
(B)(4). PRODUCT DOES NOT RETURNED FOR EVALUATION YET. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER'S TEST STRIP HAD A POOR STORAGE.
CUSTOMER COMPLAINS OF LOW RESULTS. CUSTOMER STATES THAT HE FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. THE CUSTOMER'S EXPECTED BLOOD RESULT IS 120-160MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 10/21/2018. CUSTOMER CONFIRMS THE STRIPS HAVE BEEN STORED IN THE KITCHEN AND WERE FIRST OPENED (B)(6) 2015. CUSTOMER PERFORMED A BACK TO BACK BLOOD TEST AND OBTAINED 65MG/DL AND 69MG/DL FASTING. REVIEWED METER MEMORY 78MG/DL, (B)(6) 2016 04:33:00 PM, FASTING:YES. 65MG/DL, (B)(6) 2016 02:59:00 PM, FASTING:YES. 58MG/DL, (B)(6) 2016 11:00:00 AM, FASTING:YES. 150MG/DL, (B)(6) 2016 04:04:00 PM, FASTING:YES. 138MG/DL, (B)(6) 2016 12:44:00 PM, FASTING:YES. MEMORY CONCERNS:78,65,58MG/DL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 100370 | TRUERESULT | BLOOD GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | TRUERESULT | PS2542 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |