CAIMAN DISP.INSTR.NON ARTICUL.D:5/360MM
Report
- Report Number
- 2916714-2016-00082
- Event Type
- Malfunction
- Date Received
- February 18, 2016
- Date of Event
- December 14, 2015
- Report Date
- May 24, 2016
- Manufacturer
- AESCULAP AG
- Product Code
- GEI
- PMA / PMN Number
- K130596
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS RETURNED IN A CONTAMINATED STATE. THERE IS A CLEAR VISIBLE CHARRING AT THE LEFT SIDE OF THE UPPER JAW. VISIBLE INSPECTION OF THE JAW INDICATED BLOOD SOILING AND CHARRING, HOWEVER, NO OTHER ABNORMALITIES WERE NOTED. MECHANICAL FUNCTION TESTING OF THE RECEIVED DEVICE WAS COMPLETED AND THE CLAMPING AND THE CUTTING FUNCTION WORKED WELL. THE MANUFACTURING DOCUMENTS WERE REVIEWED AND FOUND TO BE ACCORDING TO SPECIFICATION VALID DURING THE TIME OF PRODUCTION. THE MOST PROBABLY ROOT CAUSE IN CASES LIKE THIS IS A IMPROPER CLEANING DURING SURGERY; RESPECTIVELY CLEANING WITH SALINE SOLUTION, SO THAT CARBONIZED RESIDUES OF BLOOD OR TISSUE IN COMMON WITH SALT INITIATE AN ARC. A FURTHER POSSIBLE ROOT CAUSE IS A DAMAGED INSULATION TONGUE (DISSECTION CLIP). THIS DAMAGE IS CREATED BY INCORRECT HANDLING DURING CLEANING THE ELECTRODES. CORRECTIVE / PREVENTIVE ACTION: NOT REQUIRED.
PRIOR 510(K): K110824.
DURING A TOTAL LAPAROSCOPIC HYSTERECTOMY THE DEVICE SPARKED INSIDE THE PATIENT. DEVICE WAS REMOVED, THERE WAS NO INJURY TO THE PATIENT. SURGICAL DELAY OF APPROXIMATELY 5 MINUTES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 102379 | CAIMAN DISP.INSTR.NON ARTICUL.D:5/360MM | CUT AND SEAL DEVICE | GEI | AESCULAP AG | PL720SU | 52177312 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |